Drugs & medicalClass I

DATEX--OHMEDA, INC. recalls GE Heathcare Giraffe OmniBed; neo natal incubator

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This…

Read the official notice →
Date
2024-10-23
Country
United States
Brand
Datex-Ohmeda
Product
GE Heathcare Giraffe OmniBed; neo natal incubator
Classification
Class I
Units
Worldwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Datex-Ohmeda recalls

  1. 2026-09-23
  2. US
    Drugs & medicalClass II

    Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

    Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling…

    FDA · DeviceDatex-Ohmeda, Inc.Worldwide - US Nationwide distribution.Official notice →
  3. 2026-06-24
  4. US
    Drugs & medicalClass I

    Datex-Ohmeda Inc recalls Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001,…

    Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting…

    FDA · DeviceDatex-Ohmeda IncPendingOfficial notice →
  5. 2024-11-06
  6. US
    Drugs & medicalClass I

    DATEX--OHMEDA, INC. recalls GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3)…

    Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week.…

    FDA · DeviceDATEX--OHMEDA, INC.worldwideOfficial notice →
  7. 2024-10-30
  8. US
    Drugs & medicalClass I

    Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to…

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    FDA · DeviceDatex-Ohmeda, Inc.Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →
  9. 2024-02-07
  10. US
    Drugs & medicalClass I

    Datex-Ohmeda, Inc. recalls EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

    GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These…

    FDA · DeviceDatex-Ohmeda, Inc.WorldwideOfficial notice →
  11. 2023-07-12
  12. US
    Drugs & medicalClass II

    DATEX--OHMEDA, INC. recalls Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f.…

    GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe…

    FDA · DeviceDATEX--OHMEDA, INC.Worldwide distribution.Official notice →
All 7 Datex-Ohmeda recalls →

Other drugs & medical recalls in the United States

  1. 2024-10-23
  2. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls Melrose Wakefield Hospital REF K05T-2159 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe…

    Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.

    FDA · DeviceMerit Medical Systems, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  3. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light…

    Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass I

    Kinova Inc recalls KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. Jaco replaces the…

    Potential installation issue that could result in a fire hazard in the event of an interaction between a damaged Jaco arm and a wheelchair exhibiting electrical leakage.

    FDA · DeviceKinova IncWorldwide - US Nationwide distribution in the states of California, Missouri,…Official notice →
  5. US
    Drugs & medicalClass II

    Wilson-Cook Medical Inc. recalls Cook Medical Hemo-7 Hemospray Endoscopic Hemostat, REF G56572, 2.8 mm, Sterile; and Hemo-10 Hemospray Endoscopic Hemostat, REF G21049,…

    Specific lots of product have been manufactured with nonconforming raw materials.

    FDA · DeviceWilson-Cook Medical Inc.Distribution was nationwide. There was government and military distribution.…Official notice →
  6. US
    Drugs & medicalClass II

    Zhejiang Jingwei Pharmaceutical Co., Ltd. recalls First Aid Antiseptic Ointment, Povidone Iodine USP 10%, Inactive Ingredients: mineral oil, petrolatum. NET Wt. 1 OZ…

    Incorrect/Undeclared Excipients: The inactive ingredients labeled on the product boxes and tubes are listed as "mineral oil , petrolatum ". The inactive ingredients in the actual product are polyethylene glycol 400 ,…

    FDA · DrugZhejiang Jingwei Pharmaceutical Co., Ltd.Distributed to one wholesaler in OH for Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass I

    Bionpharma Inc. recalls Atovaquone Oral Suspension, USP 750 mg/5 mL, 210 mL bottle, Rx only, Distributed by: Bionpharma Inc., Princeton, NJ 08540, NDC #…

    Microbial Contamination of a Non-sterile Product: The product was found to be contaminated with Cohnella bacteria.

    FDA · DrugBionpharma Inc.Nationwide in the U.S.A.Official notice →