Datex-Ohmeda recalls

7 official recalls and safety alerts involving Datex-Ohmeda since 2023-07, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Datex-Ohmeda or the relevant agency.

  1. 2026-09-23
  2. US
    Drugs & medicalClass II

    Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

    Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling…

    FDA · DeviceDatex-Ohmeda, Inc.Worldwide - US Nationwide distribution.Official notice →
  3. 2026-06-24
  4. US
    Drugs & medicalClass I

    Datex-Ohmeda Inc recalls Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001,…

    Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting…

    FDA · DeviceDatex-Ohmeda IncPendingOfficial notice →
  5. 2024-11-06
  6. US
    Drugs & medicalClass I

    DATEX--OHMEDA, INC. recalls GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3)…

    Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week.…

    FDA · DeviceDATEX--OHMEDA, INC.worldwideOfficial notice →
  7. 2024-10-30
  8. US
    Drugs & medicalClass I

    Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to…

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    FDA · DeviceDatex-Ohmeda, Inc.Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →
  9. 2024-10-23
  10. US
    Drugs & medicalClass I

    DATEX--OHMEDA, INC. recalls GE Heathcare Giraffe OmniBed; neo natal incubator

    GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This…

    FDA · DeviceDATEX--OHMEDA, INC.Worldwide distribution.Official notice →
  11. 2024-02-07
  12. US
    Drugs & medicalClass I

    Datex-Ohmeda, Inc. recalls EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

    GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These…

    FDA · DeviceDatex-Ohmeda, Inc.WorldwideOfficial notice →
  13. 2023-07-12
  14. US
    Drugs & medicalClass II

    DATEX--OHMEDA, INC. recalls Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f.…

    GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe…

    FDA · DeviceDATEX--OHMEDA, INC.Worldwide distribution.Official notice →