Drugs & medicalClass II

Ethicon Endo-Surgery Inc recalls ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload…

There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may…

Read the official notice →
Date
2024-02-07
Country
United States
Brand
Ethicon Endo-Surgery
Product
ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping…
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Canada and Brazil.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Ethicon Endo-Surgery recalls

  1. 2025-10-22
  2. US
    Drugs & medicalClass II

    Ethicon Endo-Surgery Inc recalls ECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection, and/or creation of anastomoses, Product Number GLC80

    The devices were shipped unsterilized.

    FDA · DeviceEthicon Endo-Surgery IncInternational distribution to the countries of United Arab Emirates.Official notice →
  3. 2025-05-28
  4. US
    Drugs & medicalClass I

    Ethicon Endo-Surgery Inc recalls Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35

    Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple…

    FDA · DeviceEthicon Endo-Surgery IncUS distribution nationwide. OUS distribution to EMEA, LATAM, APAC, and Canada.Official notice →
  5. 2025-04-09
  6. US
    Drugs & medicalClass II

    Ethicon Endo Surgery, LLC recalls PDS¿PLUS¿VIO¿30IN(75CM)¿USP6-0(M0.7)¿D/A¿C-1¿MP

    Identified curing issues with the silicone during the needles manufacturing process.

    FDA · DeviceEthicon Endo Surgery, LLCWorldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of:…Official notice →
All 4 Ethicon Endo-Surgery recalls →

Other drugs & medical recalls in the United States

  1. 2024-02-07
  2. US
    Drugs & medicalClass II

    FX SHOULDER recalls Centered Glenosphere w/Screw, CoCr/TA6V/TiN 10-degrees Tilt 32 mm, REF: 115-3210

    Centered glenosphere box may contain an eccentric glenosphere.

    FDA · DeviceFX SHOULDERUS Nationwide distribution in the state of NJ.Official notice →
  3. US
    Drugs & medicalClass II

    TERRAGENE S.A. recalls Terragene Bionova PCD (PCD222-C)

    Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products.

    FDA · DeviceTERRAGENE S.A.U.S. Nationwide distribution in the states of GA, IN, MI, MT, NY, PA, TX and VA.Official notice →
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12…

    System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    American Contract Systems, Inc. recalls Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE…

    During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked"…

    FDA · DeviceAmerican Contract Systems, Inc.US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass I

    Datex-Ohmeda, Inc. recalls EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

    GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These…

    FDA · DeviceDatex-Ohmeda, Inc.WorldwideOfficial notice →
  7. US
See all drugs & medical recalls →