Drugs & medicalClass II

CooperSurgical, Inc. recalls Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark II Benchtop Incubator (AY200246)

There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an…

Read the official notice →
Date
2022-10-05
Country
United States
Brand
CooperSurgical
Product
Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark II Benchtop Incubator (AY200246)
Classification
Class II
Units
Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More CooperSurgical recalls

  1. 2023-04-12
  2. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060

    It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and…

    FDA · DeviceCooperSurgical, Inc.Worldwide - US Nationwide distribution in the states of NY, UT and the…Official notice →
  3. 2023-02-08
  4. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls BIOPSY MEDIUM 10ML, Reference Number 10620010 A, used for blastomere biopsy of cleavage stage embryos for Pre-implantation Genetic…

    The affected product may contain a medium other than the Biopsy Medium. There is a risk that use of the Product from this lot could cause degradation of the embryo.

    FDA · DeviceCooperSurgical, Inc.Global Distribution including countries of: Bahrain, Belgium, Bulgaria, Egypt,…Official notice →
  5. 2022-11-16
  6. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540

    Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In…

    FDA · DeviceCooperSurgical, Inc.Worldwide distribution: US (Nationwide) including Puerto Rico; and countries…Official notice →
  7. 2022-07-27
  8. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055

    There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English…

    FDA · DeviceCooperSurgical, Inc.Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT,…Official notice →
  9. 2022-05-25
  10. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls VITRIGUARD, CARRIER, CLEAR; Model. No. 43782001 The VitriGuard device is a cryopreservation storage device composed of a stick and cap.

    Products in this lot may have compromised barrier seals, introducing a risk to sterility.

    FDA · DeviceCooperSurgical, Inc.Domestic distribution to AZ, CA, FL, IL, MI, MO, NY, OR, TX, and WV. No…Official notice →
  11. 2022-02-23
  12. US
All 17 CooperSurgical recalls →

Other drugs & medical recalls in the United States

  1. 2022-10-05
  2. US
    Drugs & medicalClass II

    Surgical Innovations Ltd recalls Product/Part: YelloPort Plus 5x70 Cannula + Luer/YC0507001. For use in laparoscopic procedures.

    Laparoscopic devices have a sterile symbol included on labels but devices are non-sterile.

    FDA · DeviceSurgical Innovations LtdUS: MAOfficial notice →
  3. US
    Drugs & medicalClass II

    Leica Microsystems, Inc. recalls Leica Surgical Operating Microscopes, M530 OHX, Part Number 10448737

    During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the…

    FDA · DeviceLeica Microsystems, Inc.US Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Cook Incorporated recalls Fixed Core Wire Guide (Safe-T-J Curved), Guidewire Reference Part Number/GPN: TSCF-25-180-3 G01602 TSCF-35-180-3 G00902 TSCF-21-260-3…

    Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

    FDA · DeviceCook IncorporatedDomestic distribution nationwide. Foreign distribution worldwide.Official notice →
  5. US
    Drugs & medicalClass II

    Ossur Americas recalls OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA01001L OSSUR Power Knee REF PKA01001T

    Battery may dislodge from the Power Knee

    FDA · DeviceOssur AmericasU.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IA,…Official notice →
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 and IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) Siemens Material Number (SMN): 10876421 US;…

    An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results.…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide Distribution: US (nationwide) to states of: FL IN MA MI NH NJ NY OH…Official notice →
  7. US
    Drugs & medicalClass II

    Angiodynamics, Inc. recalls VenaCure EVLT KIT WITH SPOTLIGHT OPS SHEATH 55CM Ref./ Catalog No.: EVLT55OPS (each box contains 5 Procedure Kits). For use in the…

    OPS Fiber packaged within specific VenaCure EVLT Procedure Kits does not meet the required dimensional specification, may result in damage to the sheath, including separation or the inability to move the fiber within…

    FDA · DeviceAngiodynamics, Inc.US Nationwide distribution in the states of FL, NC, NJ, NM, NY, PA, TX, VA.Official notice →
See all drugs & medical recalls →