Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 and IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) Siemens Material Number (SMN): 10876421 US;…

An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results.…

Read the official notice →
Date
2022-10-05
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
IMMULITE 2000 and IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) Siemens Material Number (SMN): 10876421 US; 10713448 (OUS)
Classification
Class II
Units
Worldwide Distribution: US (nationwide) to states of: FL IN MA MI NH NJ NY OH…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2022-10-26
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Lithium (Li), Siemens Material Number 11097535. in vitro diagnostic assay

    Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →
  3. US
    Drugs & medicalClass II

    SIEMENS HEALTHCARE DIAGNOSTICS INC recalls N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297

    N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

    FDA · DeviceSIEMENS HEALTHCARE DIAGNOSTICS INCNationwideOfficial notice →
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN):…

    Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: CanadaOfficial notice →
  5. 2022-10-05
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57

    Dimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwided Foreign: Afghanistan¿¿¿¿ Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿…Official notice →
  7. 2022-09-07
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3.…

    Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Austria,…Official notice →
  9. 2022-08-10
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and…

    Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS,…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2022-10-05
  2. US
    Drugs & medicalClass II

    Cook Incorporated recalls Fixed Core Wire Guide (Safe-T-J Curved), Guidewire Reference Part Number/GPN: TSCF-25-180-3 G01602 TSCF-35-180-3 G00902 TSCF-21-260-3…

    Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

    FDA · DeviceCook IncorporatedDomestic distribution nationwide. Foreign distribution worldwide.Official notice →
  3. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark II Benchtop Incubator (AY200246)

    There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an…

    FDA · DeviceCooperSurgical, Inc.Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS…Official notice →
  4. US
    Drugs & medicalClass II

    Ossur Americas recalls OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA01001L OSSUR Power Knee REF PKA01001T

    Battery may dislodge from the Power Knee

    FDA · DeviceOssur AmericasU.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IA,…Official notice →
  5. US
    Drugs & medicalClass II

    Angiodynamics, Inc. recalls VenaCure EVLT KIT WITH SPOTLIGHT OPS SHEATH 55CM Ref./ Catalog No.: EVLT55OPS (each box contains 5 Procedure Kits). For use in the…

    OPS Fiber packaged within specific VenaCure EVLT Procedure Kits does not meet the required dimensional specification, may result in damage to the sheath, including separation or the inability to move the fiber within…

    FDA · DeviceAngiodynamics, Inc.US Nationwide distribution in the states of FL, NC, NJ, NM, NY, PA, TX, VA.Official notice →
  6. US
    Drugs & medicalClass II

    DICOM Grid, Inc. d/b/a Ambra Health recalls Ambra PACS, Ambra ProViewer. Software for use as a primary diagnostic and analysis tool for diagnostic images.

    A race condition between the storage system and services database has the potential to revert edits made to patient information upon first ingestion of a study.

    FDA · DeviceDICOM Grid, Inc. d/b/a Ambra HealthDomestic: AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA,…Official notice →
  7. US
    Drugs & medicalClass III

    Baxter Healthcare Corporation recalls Starling Monitor, Product code CMMST5 and Starling Monitor SV, Product code CMM-ST5 (same products); and accessories/spare parts…

    Labeling has incorrect or incomplete Unique Device Identification (UDI) information; and the barcodes are not compliant with GS-1 standards.

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide Distribution and government/military distribution.Official notice →
See all drugs & medical recalls →