Drugs & medicalClass II

CooperSurgical, Inc. recalls Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060

It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and…

Read the official notice →
Date
2023-04-12
Country
United States
Brand
CooperSurgical
Product
Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of NY, UT and the…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More CooperSurgical recalls

  1. 2024-02-21
  2. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls global Medium, Model Numbers LGGG-020 (20mL bottle), LGGG-050 (50mL bottle), LGGG-100 (100mL bottle). Intended use: Culture of human…

    The firm has become aware of a sudden increase in complaints for three associated lots. Performance issues may lead to impaired embryo development prior to the blastocyst stage.

    FDA · DeviceCooperSurgical, Inc.Worldwide - US Nationwide including in the states of AL, AZ, CA, CO, FL, GA,…Official notice →
  3. 2024-02-14
  4. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Box as labeled: Regard, Carter-Thomason CloseSure System XL, REF 733776, Quantity 1 Box of 3 Systems. Kit as labeled: Regard,…

    Incorrect component descriptions within the "Contents" section on the Tyvek lid for individual kit trays.

    FDA · DeviceCooperSurgical, Inc.US Nationwide distribution in the state of MO.Official notice →
  5. 2023-12-13
  6. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Cooper Surgical H/S ELLIPTOSPHERE CATH 5 Fr -For administering contrast media during Hysterosalpingography or Hysterosonography procedures…

    Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure

    FDA · DeviceCooperSurgical, Inc.Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLANDOfficial notice →
  7. 2023-02-08
  8. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls BIOPSY MEDIUM 10ML, Reference Number 10620010 A, used for blastomere biopsy of cleavage stage embryos for Pre-implantation Genetic…

    The affected product may contain a medium other than the Biopsy Medium. There is a risk that use of the Product from this lot could cause degradation of the embryo.

    FDA · DeviceCooperSurgical, Inc.Global Distribution including countries of: Bahrain, Belgium, Bulgaria, Egypt,…Official notice →
  9. 2022-11-16
  10. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540

    Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In…

    FDA · DeviceCooperSurgical, Inc.Worldwide distribution: US (Nationwide) including Puerto Rico; and countries…Official notice →
  11. 2022-10-05
  12. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark II Benchtop Incubator (AY200246)

    There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an…

    FDA · DeviceCooperSurgical, Inc.Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS…Official notice →
All 17 CooperSurgical recalls →

Other drugs & medical recalls in the United States

  1. 2023-04-19
  2. US
  3. US
    Drugs & medicalClass III

    SUN PHARMACEUTICAL INDUSTRIES INC recalls Norepinephrine Bitartrate Injection, USP, 4 mg/4 mL* (1 mg/mL), 4 mL Single-dose Fliptop Vial (NDC 47335-615-40); packaged in…

    Failed Impurities/Degradation Specifications: Above the specification limits yielded for related substance norepinephrine sulfonic acid impurity during routine product monitoring.

    FDA · DrugSUN PHARMACEUTICAL INDUSTRIES INCNationwide in the USAOfficial notice →
  4. 2023-04-12
  5. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

    The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This…

    FDA · DeviceAbbott Molecular, Inc.US Distribution to state of: PA, a government account.Official notice →
  6. US
    Drugs & medicalClass II

    Terumo Cardiovascular Systems Corporation recalls CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE

    Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly…

    FDA · DeviceTerumo Cardiovascular Systems CorporationUS Nationwide. Belgium, Canada, Costa Rica, Thailand.Official notice →
  7. US
    Drugs & medicalClass II

    Stryker Corporation recalls Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,1PRT QUICK Catalog number: 5921-030-135NS Indicated for use…

    Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood…

    FDA · DeviceStryker CorporationUS Nationwide distribution.Official notice →
  8. US
    Drugs & medicalClass II

    Moberg Research, Inc. recalls Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and…

    Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC

    FDA · DeviceMoberg Research, Inc.Worldwide distribution - US Nationwide and the countries of France, Germany,…Official notice →