Drugs & medicalClass II

Abbott Molecular, Inc. recalls Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This…

Read the official notice →
Date
2023-04-12
Country
United States
Brand
Abbott Molecular
Product
Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.
Classification
Class II
Units
US Distribution to state of: PA, a government account.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Molecular recalls

  1. 2024-05-01
  2. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System, Part Number: 08N53-002

    The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA,…Official notice →
  3. 2024-03-06
  4. US
    Drugs & medicalClass III

    Abbott Molecular, Inc. recalls Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)

    Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.

    FDA · DeviceAbbott Molecular, Inc.US Nationwide distribution in the states of AL, CA, IL, MA, MD, MN, NY, and PA.Official notice →
  5. 2023-12-06
  6. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse…

    Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

    FDA · DeviceAbbott Molecular, Inc.WorldwideOfficial notice →
  7. 2023-02-01
  8. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing,…

    The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration…

    FDA · DeviceAbbott Molecular, Inc.xOfficial notice →
  9. 2022-12-28
  10. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m Resp-4-Plex Amp Kit, Part Number 09N79-096

    Reports of an increase in reactive negative controls and false positive results with certain lots of Alinity M Resp-4-Plex AMP kits.

    FDA · DeviceAbbott Molecular, Inc.WorldwideOfficial notice →
  11. 2022-11-30
  12. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

    Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution.Official notice →
All 21 Abbott Molecular recalls →

Other drugs & medical recalls in the United States

  1. 2023-04-19
  2. US
    Drugs & medicalClass III

    Pine Pharmaceuticals, LLC recalls Bevacizumab 2.5 mg/0.1 mL, Solution for Injection in 1mL, silicone free slip tip syringe, each Syringe supplied in individually labeled…

    Labeling: Incorrect or Missing Lot and/or Exp Date: Lot code on primary packaging is incorrect.

    FDA · DrugPine Pharmaceuticals, LLCNationwide in the USAOfficial notice →
  3. US
  4. US
    Drugs & medicalClass III

    SUN PHARMACEUTICAL INDUSTRIES INC recalls Norepinephrine Bitartrate Injection, USP, 4 mg/4 mL* (1 mg/mL), 4 mL Single-dose Fliptop Vial (NDC 47335-615-40); packaged in…

    Failed Impurities/Degradation Specifications: Above the specification limits yielded for related substance norepinephrine sulfonic acid impurity during routine product monitoring.

    FDA · DrugSUN PHARMACEUTICAL INDUSTRIES INCNationwide in the USAOfficial notice →
  5. 2023-04-12
  6. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060

    It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and…

    FDA · DeviceCooperSurgical, Inc.Worldwide - US Nationwide distribution in the states of NY, UT and the…Official notice →
  7. US
    Drugs & medicalClass II

    Terumo Cardiovascular Systems Corporation recalls CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE

    Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly…

    FDA · DeviceTerumo Cardiovascular Systems CorporationUS Nationwide. Belgium, Canada, Costa Rica, Thailand.Official notice →
  8. US
    Drugs & medicalClass II

    Stryker Corporation recalls Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,1PRT QUICK Catalog number: 5921-030-135NS Indicated for use…

    Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood…

    FDA · DeviceStryker CorporationUS Nationwide distribution.Official notice →
See all drugs & medical recalls →