Drugs & medicalClass II

Abbott Molecular, Inc. recalls Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse…

Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

Read the official notice →
Date
2023-12-06
Country
United States
Brand
Abbott Molecular
Product
Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse…
Classification
Class II
Units
Worldwide
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Molecular recalls

  1. 2024-11-06
  2. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091

    Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution: US (nationwide) to states of: AZ, CA, CO, DC, FL, GA,…Official notice →
  3. 2024-05-01
  4. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System, Part Number: 08N53-002

    The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA,…Official notice →
  5. 2024-03-06
  6. US
    Drugs & medicalClass III

    Abbott Molecular, Inc. recalls Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)

    Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.

    FDA · DeviceAbbott Molecular, Inc.US Nationwide distribution in the states of AL, CA, IL, MA, MD, MN, NY, and PA.Official notice →
  7. 2023-04-12
  8. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

    The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This…

    FDA · DeviceAbbott Molecular, Inc.US Distribution to state of: PA, a government account.Official notice →
  9. 2023-02-01
  10. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing,…

    The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration…

    FDA · DeviceAbbott Molecular, Inc.xOfficial notice →
  11. 2022-12-28
  12. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m Resp-4-Plex Amp Kit, Part Number 09N79-096

    Reports of an increase in reactive negative controls and false positive results with certain lots of Alinity M Resp-4-Plex AMP kits.

    FDA · DeviceAbbott Molecular, Inc.WorldwideOfficial notice →
All 21 Abbott Molecular recalls →

Other drugs & medical recalls in the United States

  1. 2023-12-06
  2. US
    Drugs & medicalClass II

    Aizu Olympus Co., Ltd. recalls PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE

    Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or…

    FDA · DeviceAizu Olympus Co., Ltd.Domestic distribution to FL, KY, MA, NJ, NY, and PAOfficial notice →
  3. US
    Drugs & medicalClass II

    Edwards Lifesciences, LLC recalls Acumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.

    Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.

    FDA · DeviceEdwards Lifesciences, LLCUS: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO…Official notice →
  4. US
    Drugs & medicalClass II

    EMD Millipore Corporation recalls Lugol s Iodine Solution 500ML, Part Number 624-71

    Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop…

    FDA · DeviceEMD Millipore CorporationUS States: CA, NJ, WVOfficial notice →
  5. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Da Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mm

    There is the potential for the needle drive to become dislodged and/or broken which could result in uncontrollable movement.

    FDA · DeviceIntuitive Surgical, Inc.US: IL, FL, ID, SC, MI, TX, TN, NY, OR, CO, CA, MS, NJ, Puerto Rico OUS: South…Official notice →
  6. US
    Drugs & medicalClass II

    Breas Medical, Inc. recalls with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS.…

    Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation

    FDA · DeviceBreas Medical, Inc.Nationwide Foreign: CanadaOfficial notice →
  7. US
    Drugs & medicalClass II

    PATH recalls SENTIERO SCREENING, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098-S

    Affected lots of product may experience technical distortions to a greater extent than expected.

    FDA · DevicePATHUS States: FL, MO, OH, WIOfficial notice →