Drugs & medicalClass II

Abbott Molecular, Inc. recalls Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091

Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay…

Read the official notice →
Date
2024-11-06
Country
United States
Brand
Abbott Molecular
Product
Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091
Classification
Class II
Units
Worldwide distribution: US (nationwide) to states of: AZ, CA, CO, DC, FL, GA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Molecular recalls

  1. 2025-05-21
  2. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR)…

    There were reports of an increase in reactive negative controls and false positive results.

    FDA · DeviceAbbott Molecular, Inc.Worldwide - US Nationwide distribution.Official notice →
  3. 2025-04-30
  4. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001,…

    The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the…

    FDA · DeviceAbbott Molecular, Inc.Worldwide - US Nationwide distribution.Official notice →
  5. 2024-05-01
  6. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System, Part Number: 08N53-002

    The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA,…Official notice →
  7. 2024-03-06
  8. US
    Drugs & medicalClass III

    Abbott Molecular, Inc. recalls Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)

    Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.

    FDA · DeviceAbbott Molecular, Inc.US Nationwide distribution in the states of AL, CA, IL, MA, MD, MN, NY, and PA.Official notice →
  9. 2023-12-06
  10. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse…

    Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

    FDA · DeviceAbbott Molecular, Inc.WorldwideOfficial notice →
  11. 2023-04-12
  12. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

    The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This…

    FDA · DeviceAbbott Molecular, Inc.US Distribution to state of: PA, a government account.Official notice →
All 21 Abbott Molecular recalls →

Other drugs & medical recalls in the United States

  1. 2024-11-06
  2. US
    Drugs & medicalClass III

    DRG International, Inc. recalls 17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone…

    The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to…

    FDA · DeviceDRG International, Inc.AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy…Official notice →
  3. US
    Drugs & medicalClass II

    Covidien LP recalls Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm…

    Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.

    FDA · DeviceCovidien LPWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of:…Official notice →
  4. US
    Drugs & medicalClass II

    Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of…

    Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

    FDA · DeviceIntegra LifeSciences Corp.Worldwide Distribution. US Nationwide, Canada Germany United Kingdom of Great…Official notice →
  5. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a…

    Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the…

    FDA · DeviceOlympus Corporation of the AmericasOHOfficial notice →
  6. US
    Drugs & medicalClass II

    ImaCor Inc. recalls ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number:…

    Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

    FDA · DeviceImaCor Inc.FL ,GA, NJ, TNOfficial notice →
  7. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o…

    Due to an out of tolerance tool being used on monitor/defibrillator systems.

    FDA · DevicePhysio-Control, Inc.U.S.: CT, GA, MA, NY, OH, WA, and WV. O.U.S.: N/AOfficial notice →
See all drugs & medical recalls →