Drugs & medicalClass II

Abbott Molecular, Inc. recalls Alinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR)…

There were reports of an increase in reactive negative controls and false positive results.

Read the official notice →
Date
2025-05-21
Country
United States
Brand
Abbott Molecular
Product
Alinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR)…
Classification
Class II
Units
Worldwide - US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Molecular recalls

  1. 2025-04-30
  2. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001,…

    The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the…

    FDA · DeviceAbbott Molecular, Inc.Worldwide - US Nationwide distribution.Official notice →
  3. 2024-11-06
  4. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091

    Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution: US (nationwide) to states of: AZ, CA, CO, DC, FL, GA,…Official notice →
  5. 2024-05-01
  6. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System, Part Number: 08N53-002

    The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA,…Official notice →
  7. 2024-03-06
  8. US
    Drugs & medicalClass III

    Abbott Molecular, Inc. recalls Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)

    Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.

    FDA · DeviceAbbott Molecular, Inc.US Nationwide distribution in the states of AL, CA, IL, MA, MD, MN, NY, and PA.Official notice →
  9. 2023-12-06
  10. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse…

    Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

    FDA · DeviceAbbott Molecular, Inc.WorldwideOfficial notice →
  11. 2023-04-12
  12. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

    The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This…

    FDA · DeviceAbbott Molecular, Inc.US Distribution to state of: PA, a government account.Official notice →
All 21 Abbott Molecular recalls →

Other drugs & medical recalls in the United States

  1. 2025-05-21
  2. US
    Drugs & medicalClass II

    Numed Inc recalls BIB Balloon in Balloon Dilatation Catheter. Model Number: 420

    Balloon extension lumens are not printed correctly. Specifically, the inner balloon inflation hub was labeled as the outer balloon, and the outer balloon inflation hub was labeled as the inner balloon.

    FDA · DeviceNumed IncUS distribution to Pennsylvania.Official notice →
  3. US
    Drugs & medicalClass II

    Richard Wolf GmbH recalls PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with…

    The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a…

    FDA · DeviceRichard Wolf GmbHUS State: GAOfficial notice →
  4. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56

    During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).

    FDA · DeviceLeMaitre Vascular, Inc.Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA,…Official notice →
  5. US
    Drugs & medicalClass II

    Foundation Medicine, Inc. recalls FoundationOne Companion Diagnostic (F1CDx)

    Four (4) total reports were sent to customers with clinically relevant variants requiring amendment, one (1) of which with an incorrect negative CDx claim. The remaining three (3) with tumor profiling result changes…

    FDA · DeviceFoundation Medicine, Inc.Worldwide - US Nationwide distribution in the states of GA, NY and the country…Official notice →
  6. US
    Drugs & medicalClass II

    Meddev Corp recalls Brand Name: Contour Product Name: Contour Gold Eyelid Weight Model/Catalog Number: LL3012 Product Description: Contour Gold Eyelid Weight,…

    Due to mislabeling of products

    FDA · DeviceMeddev CorpU.S. Nationwide distribution in the states of CA, CO, CT, FL, IA, IL, MA, ME,…Official notice →
  7. US
    Drugs & medicalClass II

    AVID Medical, Inc. recalls Halyard CHRIST SUTURE KIT. General Procedure Tray Convenience Kit

    An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.

    FDA · DeviceAVID Medical, Inc.US Nationwide distribution in the states of IL & LA.Official notice →