Drugs & medicalClass II

LeMaitre Vascular, Inc. recalls Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56

During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).

Read the official notice →
Date
2025-05-21
Country
United States
Brand
LeMaitre Vascular
Product
Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56
Classification
Class II
Units
Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More LeMaitre Vascular recalls

  1. 2025-10-29
  2. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls Artegraft¿ Collagen Vascular Graft; REF: AG1015;

    The device was incorrectly packed in the wrong size labeled outer packaging.

    FDA · DeviceLeMaitre Vascular, Inc.US distribution to the state of: AROfficial notice →
  3. 2025-10-15
  4. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls Artegraft Collagen Vascular Graft. Model/Catalog/Part Numbers: AG630M, AG636M, AG730M, AG740M, AG845M. The Artegraft is intended to serve…

    Devices were distributed with incorrect label without CE and UKCA mark, missing patient leaflet and patient implant card.

    FDA · DeviceLeMaitre Vascular, Inc.No US distribution. International distribution to Great Britian and Switzerland.Official notice →
  5. 2025-10-08
  6. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8)…

    The bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been reviewed and approved by the appropriate regulatory authority.

    FDA · DeviceLeMaitre Vascular, Inc.US Nationwide distribution in the states of CA, CT, FL, GA, IL, MA, MS, NC,…Official notice →
  7. 2024-05-22
  8. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.

    The guide tip can become damaged and result in the tip detaching.

    FDA · DeviceLeMaitre Vascular, Inc.Worldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →
  9. 2023-09-06
  10. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ or non-reversed…

    Burrs (molding flash) on the cutter may tear the harvested vien

    FDA · DeviceLeMaitre Vascular, Inc.International distribution to the countries of Belgium, France, Germany, and…Official notice →
  11. 2022-11-23
  12. US
    Drugs & medicalClass II

    LeMaitre Vascular, Inc. recalls TufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire…

    Three lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole,…

    FDA · DeviceLeMaitre Vascular, Inc.No US distribution. Foreign distribution to United Kingdom, Germany, France,…Official notice →
All 9 LeMaitre Vascular recalls →

Other drugs & medical recalls in the United States

  1. 2025-05-21
  2. US
    Drugs & medicalClass II

    Augmedics Ltd. recalls Augmedics Perc Pin Adaptor, Model/Catalog Number AMCH07200, a component of the xvision Spine (XVS) system, an image-guided navigation…

    The perc pin adaptor AMCH07200 tightening knob was found to be inconvenient to the user and the tightening markings found to be improperly placed.

    FDA · DeviceAugmedics Ltd.US Nationwide.Official notice →
  3. US
    Drugs & medicalClass II

    Numed Inc recalls BIB Balloon in Balloon Dilatation Catheter. Model Number: 420

    Balloon extension lumens are not printed correctly. Specifically, the inner balloon inflation hub was labeled as the outer balloon, and the outer balloon inflation hub was labeled as the inner balloon.

    FDA · DeviceNumed IncUS distribution to Pennsylvania.Official notice →
  4. US
    Drugs & medicalClass II

    Richard Wolf GmbH recalls PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with…

    The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a…

    FDA · DeviceRichard Wolf GmbHUS State: GAOfficial notice →
  5. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR)…

    There were reports of an increase in reactive negative controls and false positive results.

    FDA · DeviceAbbott Molecular, Inc.Worldwide - US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Foundation Medicine, Inc. recalls FoundationOne Companion Diagnostic (F1CDx)

    Four (4) total reports were sent to customers with clinically relevant variants requiring amendment, one (1) of which with an incorrect negative CDx claim. The remaining three (3) with tumor profiling result changes…

    FDA · DeviceFoundation Medicine, Inc.Worldwide - US Nationwide distribution in the states of GA, NY and the country…Official notice →
  7. US
    Drugs & medicalClass II

    Meddev Corp recalls Brand Name: Contour Product Name: Contour Gold Eyelid Weight Model/Catalog Number: LL3012 Product Description: Contour Gold Eyelid Weight,…

    Due to mislabeling of products

    FDA · DeviceMeddev CorpU.S. Nationwide distribution in the states of CA, CO, CT, FL, IA, IL, MA, ME,…Official notice →