LeMaitre Vascular, Inc. recalls Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.
The guide tip can become damaged and result in the tip detaching.
Read the official notice →- Date
- 2024-05-22
- Country
- United States
- Brand
- LeMaitre Vascular
- Product
- Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Australia, Canada,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More LeMaitre Vascular recalls
- 2025-10-29
US
LeMaitre Vascular, Inc. recalls Artegraft¿ Collagen Vascular Graft; REF: AG1015;
The device was incorrectly packed in the wrong size labeled outer packaging.
- 2025-10-15
US
LeMaitre Vascular, Inc. recalls Artegraft Collagen Vascular Graft. Model/Catalog/Part Numbers: AG630M, AG636M, AG730M, AG740M, AG845M. The Artegraft is intended to serve…
Devices were distributed with incorrect label without CE and UKCA mark, missing patient leaflet and patient implant card.
FDA · DeviceLeMaitre Vascular, Inc.No US distribution. International distribution to Great Britian and Switzerland.Official notice →- 2025-10-08
US
LeMaitre Vascular, Inc. recalls Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8)…
The bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been reviewed and approved by the appropriate regulatory authority.
FDA · DeviceLeMaitre Vascular, Inc.US Nationwide distribution in the states of CA, CT, FL, GA, IL, MA, MS, NC,…Official notice →- 2025-05-21
US
LeMaitre Vascular, Inc. recalls Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
FDA · DeviceLeMaitre Vascular, Inc.Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA,…Official notice →- 2023-09-06
US
LeMaitre Vascular, Inc. recalls The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ or non-reversed…
Burrs (molding flash) on the cutter may tear the harvested vien
FDA · DeviceLeMaitre Vascular, Inc.International distribution to the countries of Belgium, France, Germany, and…Official notice →- 2022-11-23
US
LeMaitre Vascular, Inc. recalls TufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire…
Three lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole,…
FDA · DeviceLeMaitre Vascular, Inc.No US distribution. Foreign distribution to United Kingdom, Germany, France,…Official notice →
Other drugs & medical recalls in the United States
- 2024-05-22
US
Andover Healthcare Inc. recalls Sterile CoFlex NL: Latex Free Flexible Cohesive Bandage, Tan, 4INx5YD (5400S)
A packaging seal gap could impact the sterility of the product.
FDA · DeviceAndover Healthcare Inc.US Nationwide distribution in the states of CA, CO, GA, IA, IN, OH, MN, NC,…Official notice →US
Philips North America Llc recalls Patient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.
Event Catalog information does not save when copied and transferred from one unit to another.
FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution in the states of CA, FL, GA, IA, MA,…Official notice →US
Sophysa S.A. recalls The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular…
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient.…
FDA · DeviceSophysa S.A.Worldwide - US Nationwide distribution in the states of AZ, NY, TX and the…Official notice →US
Fisher & Paykel Healthcare, Ltd. recalls Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound…
FDA · DeviceFisher & Paykel Healthcare, Ltd.US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →US
Fresenius Medical Care Holdings, Inc. recalls Novalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respiratory/…
Issue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during…
US
Medartis Inc. recalls APTUS 2.8 TriLock Screw 22mm, HD7, 1/Pkg. Metal bone fixation fastener used with the APTUS fixation system.
Screw is 16mm long instead of 22mm long.
FDA · DeviceMedartis Inc.Worldwide - US Nationwide distribution in the state of TX and the countries of…Official notice →