Philips North America Llc recalls Patient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.
Event Catalog information does not save when copied and transferred from one unit to another.
Read the official notice →- Date
- 2024-05-22
- Country
- United States
- Brand
- Philips
- Product
- Patient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, MA,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2024-05-29
US
Philips North America Llc recalls Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728144
Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the…
US
Philips North America Llc recalls Achieva 3.0T Model Number (REF): (1) 781345; (2) 781344; (3) 781278; (4) 781277; (5) 781177
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →- 2024-05-22
US
Philips North America Llc recalls Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a…
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the…Official notice →- 2024-05-08
US
Philips North America Llc recalls The Oncology Essentials Package of the Spectral CT imaging system.
Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and…Official notice →- 2024-05-01
US
Philips North America Llc recalls Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V…
Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Argentina, Aruba,…Official notice →- 2024-04-24
US
Philips North America Llc recalls Ingenia 1.5T S, Magnetic Resonance System.
Patient support table floor plate may be incorrectly installed.
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →
Other drugs & medical recalls in the United States
- 2024-05-22
US
Maquet Medical Systems USA recalls CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
FDA · DeviceMaquet Medical Systems USAUS Nationwide distribution in the states of NJ & TX.Official notice →US
Philips North America Llc recalls Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a…
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the…Official notice →US
Andover Healthcare Inc. recalls Sterile CoFlex NL: Latex Free Flexible Cohesive Bandage, Tan, 4INx5YD (5400S)
A packaging seal gap could impact the sterility of the product.
FDA · DeviceAndover Healthcare Inc.US Nationwide distribution in the states of CA, CO, GA, IA, IN, OH, MN, NC,…Official notice →US
Sophysa S.A. recalls The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular…
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient.…
FDA · DeviceSophysa S.A.Worldwide - US Nationwide distribution in the states of AZ, NY, TX and the…Official notice →US
LeMaitre Vascular, Inc. recalls Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.
The guide tip can become damaged and result in the tip detaching.
FDA · DeviceLeMaitre Vascular, Inc.Worldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →US
Fisher & Paykel Healthcare, Ltd. recalls Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound…
FDA · DeviceFisher & Paykel Healthcare, Ltd.US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →