Intuitive Surgical, Inc. recalls Da Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mm
There is the potential for the needle drive to become dislodged and/or broken which could result in uncontrollable movement.
Read the official notice →- Date
- 2023-12-06
- Country
- United States
- Brand
- Intuitive Surgical
- Product
- Da Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mm
- Classification
- Class II
- Units
- US: IL, FL, ID, SC, MI, TX, TN, NY, OR, CO, CA, MS, NJ, Puerto Rico OUS: South…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Intuitive Surgical recalls
- 2025-01-22
US
Intuitive Surgical, Inc. recalls 8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
FDA · DeviceIntuitive Surgical, Inc.U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado,…Official notice →- 2024-09-04
US
Intuitive Surgical, Inc. recalls Ion Endoluminal System, REF: 380748-65
There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.
FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of AZ and CA.Official notice →- 2024-04-17
US
Intuitive Surgical, Inc. recalls Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098
Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries,…
FDA · DeviceIntuitive Surgical, Inc.US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV,…Official notice →- 2023-07-12
US
Intuitive Surgical, Inc. recalls da Vinci Xi/X Tip-Up Fenestrated Grasper, REF: 470347, VER: 12
Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected…
FDA · DeviceIntuitive Surgical, Inc.Worldwide - US Nationwide distribution in the states of PR, FL, MN, PA, AZ,…Official notice →- 2023-04-05
US
Intuitive Surgical, Inc. recalls da Vinci Xi Surgical System Model IS4000 da Vinci X Surgical System Model IS4200
Due to Universal Surgical Manipulator (USM) instrument carriage being loose as a result of the linear rail component not being manufactured within specifications.
FDA · DeviceIntuitive Surgical, Inc.U.S.: Cayman Islands, Dominican Republic, Puerto Rico, United States O.U.S.:…Official notice →- 2023-03-22
US
Intuitive Surgical, Inc. recalls Patient Side Cart, REF: 380620, with instrument arms, used as part of the da Vinci X Surgical System, REF: IS4200; Patient Side Cart,…
Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled…
FDA · DeviceIntuitive Surgical, Inc.Worldwide - US Nationwide distribution in the states of FL, GA, WA and the…Official notice →
Other drugs & medical recalls in the United States
- 2023-12-06
US
Edwards Lifesciences, LLC recalls Acumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
FDA · DeviceEdwards Lifesciences, LLCUS: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO…Official notice →US
EMD Millipore Corporation recalls Lugol s Iodine Solution 500ML, Part Number 624-71
Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop…
US
Abbott Molecular, Inc. recalls Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse…
Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.
US
Breas Medical, Inc. recalls with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS.…
Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
US
PATH recalls SENTIERO SCREENING, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098-S
Affected lots of product may experience technical distortions to a greater extent than expected.
US
Becton Dickinson Infusion Therapy Systems Inc. recalls BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533
Over-the-needle, intravascular catheters may have needles that do not retract properly into the barrel, which could result in contaminated needlestick injury, blood leaking from the insertion site or catheter, and…
FDA · DeviceBecton Dickinson Infusion Therapy Systems Inc.US Nationwide distribution in the states of CA, NC, MI, WA, CO, FL, NJ, NM,…Official notice →