Drugs & medicalClass II

Intuitive Surgical, Inc. recalls 8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

Read the official notice →
Date
2025-01-22
Country
United States
Brand
Intuitive Surgical
Product
8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401
Classification
Class II
Units
U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Intuitive Surgical recalls

  1. 2025-12-31
  2. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

    Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be…

    FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC,…Official notice →
  3. 2025-03-19
  4. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer…

    Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed

    FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - US Nationwide and the countries of Alaska, Arizona,…Official notice →
  5. 2025-01-22
  6. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls 8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of…

    Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments

    FDA · DeviceIntuitive Surgical, Inc.U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA,…Official notice →
  7. 2024-09-04
  8. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Ion Endoluminal System, REF: 380748-65

    There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.

    FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of AZ and CA.Official notice →
  9. 2024-04-17
  10. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

    Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries,…

    FDA · DeviceIntuitive Surgical, Inc.US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV,…Official notice →
  11. 2023-12-06
  12. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Da Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mm

    There is the potential for the needle drive to become dislodged and/or broken which could result in uncontrollable movement.

    FDA · DeviceIntuitive Surgical, Inc.US: IL, FL, ID, SC, MI, TX, TN, NY, OR, CO, CA, MS, NJ, Puerto Rico OUS: South…Official notice →
All 32 Intuitive Surgical recalls →

Other drugs & medical recalls in the United States

  1. 2025-01-22
  2. US
    Drugs & medicalClass II

    QUIDEL ORTHO recalls Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version:…

    Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.

    FDA · DeviceQUIDEL ORTHODomestic: Nationwide Distribution International: Australia, Belgium, Bermuda,…Official notice →
  3. US
    Drugs & medicalClass II

    GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomography,…

    GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices)…

    FDA · DeviceGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGINGWorldwideOfficial notice →
  4. US
    Drugs & medicalClass I

    Baxter Healthcare Corporation recalls Welch Allyn, Inc., Life 2000 Ventilation System, REF BT-20-0002

    There is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.

    FDA · DeviceBaxter Healthcare CorporationUS, Nationwide DistributionOfficial notice →
  5. US
    Drugs & medicalClass II

    Tornier S.A.S. recalls Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty.

    The Pin Guide is being recalled because the metal tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset being unable to detect the Pin Guide and potentially prolonging…

    FDA · DeviceTornier S.A.S.US: MN, KY, MD, MA, TX, AK, TX, ID; France, CanadaOfficial notice →
  6. US
    Drugs & medicalClass II

    Coloplast Manufacturing US, LLC recalls X-Flow prostatectomy catheter, Device Reference Numbers AB6022, AB6024, AB6118, AB6120, AB6122, AB6124, AB6220, AB6222, AB6224,…

    A possible sterility issue was detected in Coloplast's facility on some Coloplast products.

    FDA · DeviceColoplast Manufacturing US, LLCUS Nationwide. Canada.Official notice →
  7. US
    Drugs & medicalClass II

    Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC77V; 2)…

    HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".

    FDA · DeviceAugustine Temperature Management, LLCWorldwideOfficial notice →
See all drugs & medical recalls →