CooperSurgical, Inc. recalls LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055
There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English…
Read the official notice →- Date
- 2022-07-27
- Country
- United States
- Brand
- CooperSurgical
- Product
- LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055
- Classification
- Class II
- Units
- Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More CooperSurgical recalls
- 2023-02-08
US
CooperSurgical, Inc. recalls BIOPSY MEDIUM 10ML, Reference Number 10620010 A, used for blastomere biopsy of cleavage stage embryos for Pre-implantation Genetic…
The affected product may contain a medium other than the Biopsy Medium. There is a risk that use of the Product from this lot could cause degradation of the embryo.
FDA · DeviceCooperSurgical, Inc.Global Distribution including countries of: Bahrain, Belgium, Bulgaria, Egypt,…Official notice →- 2022-11-16
US
CooperSurgical, Inc. recalls NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540
Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In…
FDA · DeviceCooperSurgical, Inc.Worldwide distribution: US (Nationwide) including Puerto Rico; and countries…Official notice →- 2022-10-05
US
CooperSurgical, Inc. recalls Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark II Benchtop Incubator (AY200246)
There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an…
FDA · DeviceCooperSurgical, Inc.Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS…Official notice →- 2022-05-25
US
CooperSurgical, Inc. recalls VITRIGUARD, CARRIER, CLEAR; Model. No. 43782001 The VitriGuard device is a cryopreservation storage device composed of a stick and cap.
Products in this lot may have compromised barrier seals, introducing a risk to sterility.
FDA · DeviceCooperSurgical, Inc.Domestic distribution to AZ, CA, FL, IL, MI, MO, NY, OR, TX, and WV. No…Official notice →- 2022-02-23
US
- 2021-11-24
US
CooperSurgical, Inc. recalls Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number: 909075-05W
XXX
FDA · DeviceCooperSurgical, Inc.US Nationwide Distribution Foreign: ARMENIA, BOLIVIA, CAMBODIA, CANADA, CHILE,…Official notice →
Other drugs & medical recalls in the United States
- 2022-07-27
US
Siemens Medical Solutions USA, Inc recalls Sensis with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or…
Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide Distribution AZ FL KY MI OH PR TXOfficial notice →US
Randox Laboratories Ltd. recalls Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0
Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal…
US
Randox Laboratories Ltd. recalls Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in…
US
Ambu Inc. recalls Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000,…
Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.
FDA · DeviceAmbu Inc.Domestic distribution nationwide. Foreign distribution to Austria, Belgium,…Official notice →US
DePuy Orthopaedics, Inc. recalls ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant…
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical…
FDA · DeviceDePuy Orthopaedics, Inc.Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil,…Official notice →US
Smith & Nephew, Inc. recalls (1) Genesis II Non-Porous Tibial Baseplate Size 5 Right, Product #71420186. (2) Genesis II Non-Porous Tibial Baseplate Size 4 Left,…
The product was mispackaged (swapped). The devices within the inner tray matches the inner tray but does not match the outer box label and the chart sticks.
FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of AZ, CA, GA, KS, LA, MN, MI, MO,…Official notice →