Drugs & medicalClass II

Ambu Inc. recalls Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000,…

Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.

Read the official notice →
Date
2022-07-27
Country
United States
Brand
Ambu
Product
Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000,…
Classification
Class II
Units
Domestic distribution nationwide. Foreign distribution to Austria, Belgium,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Ambu recalls

  1. 2025-08-20
  2. US
    Drugs & medicalClass I

    Ambu Inc. recalls Ambu SPUR II Labeled as the following: 1 SPUR II Adult Resuscitator, Adult, Catalog Number: 523211000 2 SPUR II Adult Resuscitator…

    Potential for the manometer port being blocked rendering the manometer non-functional.

    FDA · DeviceAmbu Inc.Worldwide distribution - US Nationwide and the country of Canada.Official notice →
  3. 2025-05-14
  4. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu aScope" 5 Broncho HD 5.6/2.8 Sampler Set. Model Number: 622002000US

    Incorrect labeling in which the front red pouch label did not match the actual size of the medical device. The label on the front of the scope incorrectly states 5.0/2.2 while the back label, shipping box and product…

    FDA · DeviceAmbu Inc.US Nationwide distribution in the states of CA, CO, FL, GA, KY, MD, MT, NJ,…Official notice →
  5. 2024-11-06
  6. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number…

    Ambu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.

    FDA · DeviceAmbu Inc.US Nationwide. Canada.Official notice →
  7. 2024-01-17
  8. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

    Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the…

    FDA · DeviceAmbu Inc.Domestic distribution nationwide.Official notice →
  9. 2023-10-04
  10. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu aView 2 A Monitor-The Ambu aView 2 Advance is intended to display live imaging data from compatible Ambu visualization devices,…

    Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames

    FDA · DeviceAmbu Inc.NationwideOfficial notice →
All 6 Ambu recalls →

Other drugs & medical recalls in the United States

  1. 2022-07-27
  2. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0

    Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal…

    FDA · DeviceRandox Laboratories Ltd.Nationwide Distribution - MD, OR, PR, MT, WVOfficial notice →
  3. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD

    A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the state of IL.Official notice →
  4. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055

    There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English…

    FDA · DeviceCooperSurgical, Inc.Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT,…Official notice →
  5. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant…

    Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical…

    FDA · DeviceDePuy Orthopaedics, Inc.Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil,…Official notice →
  6. US
    Drugs & medicalClass II

    Smith & Nephew, Inc. recalls (1) Genesis II Non-Porous Tibial Baseplate Size 5 Right, Product #71420186. (2) Genesis II Non-Porous Tibial Baseplate Size 4 Left,…

    The product was mispackaged (swapped). The devices within the inner tray matches the inner tray but does not match the outer box label and the chart sticks.

    FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of AZ, CA, GA, KS, LA, MN, MI, MO,…Official notice →
  7. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress, REF P7924A01; and REF P7924ARENT01, PRO+ MRS Rental Surface with X-Ray mattress.

    The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of…

    FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →