Drugs & medicalClass II

Ambu Inc. recalls Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the…

Read the official notice →
Date
2024-01-17
Country
United States
Brand
Ambu
Product
Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2
Classification
Class II
Units
Domestic distribution nationwide.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Ambu recalls

  1. 2025-08-20
  2. US
    Drugs & medicalClass I

    Ambu Inc. recalls Ambu SPUR II Labeled as the following: 1 SPUR II Adult Resuscitator, Adult, Catalog Number: 523211000 2 SPUR II Adult Resuscitator…

    Potential for the manometer port being blocked rendering the manometer non-functional.

    FDA · DeviceAmbu Inc.Worldwide distribution - US Nationwide and the country of Canada.Official notice →
  3. 2025-05-14
  4. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu aScope" 5 Broncho HD 5.6/2.8 Sampler Set. Model Number: 622002000US

    Incorrect labeling in which the front red pouch label did not match the actual size of the medical device. The label on the front of the scope incorrectly states 5.0/2.2 while the back label, shipping box and product…

    FDA · DeviceAmbu Inc.US Nationwide distribution in the states of CA, CO, FL, GA, KY, MD, MT, NJ,…Official notice →
  5. 2024-11-06
  6. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number…

    Ambu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.

    FDA · DeviceAmbu Inc.US Nationwide. Canada.Official notice →
  7. 2023-10-04
  8. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu aView 2 A Monitor-The Ambu aView 2 Advance is intended to display live imaging data from compatible Ambu visualization devices,…

    Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames

    FDA · DeviceAmbu Inc.NationwideOfficial notice →
  9. 2022-07-27
  10. US
    Drugs & medicalClass II

    Ambu Inc. recalls Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000,…

    Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.

    FDA · DeviceAmbu Inc.Domestic distribution nationwide. Foreign distribution to Austria, Belgium,…Official notice →
All 6 Ambu recalls →

Other drugs & medical recalls in the United States

  1. 2024-01-24
  2. US
    Drugs & medicalClass II

    Par Sterile Products LLC recalls Vasostrict (vasopressin in 5% Dextrose) Injection, 20 units per 100 mL (0.2 units per mL), 100 mL x 10 Single Dose Vials per carton,…

    Superpotent Drug: Assay from the 3-month and 6-month stability intervals exceeded the upper specification limit.

    FDA · DrugPar Sterile Products LLCNationwide USAOfficial notice →
  3. US
    Drugs & medicalClass II

    Pharma Nobis LLC recalls Freskaro refresh & revitalize Magnesium Citrate Saline Laxative Oral Solution Grape Flavor 10 fl. oz, Manufactured by: Pharma Lobis, 7400…

    MICROBIAL CONTAMINATION OF NON-STERILE PRODUCTS - Presence of Acetobacter nitrogenifigens bacteria.

    FDA · DrugPharma Nobis LLCNationwide.Official notice →
  4. 2024-01-17
  5. US
    Drugs & medicalClass II

    Philips North America recalls (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)

    The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)

    FDA · DevicePhilips North AmericaDomestic distribution to the following states and territories: CA DE FL GA IL…Official notice →
  6. US
    Drugs & medicalClass II

    mo-Vis BVBA recalls Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering…

    In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the…

    FDA · Devicemo-Vis BVBAUS States: TN, TX.Official notice →
  7. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls QMS Tacrolimus Calibrators, Product code 10015573

    Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to…Official notice →
  8. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline Gown Poly W/Breath Sleeves XL AAMIK Level 4 (Eclipse gown), REF SPT-2208CS, 100/pack, Non-Sterile.

    Surgical gowns were manufactured with the wrong sleeve.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldDistribution was made to MD, OH, TN, TX, and WI. There was no…Official notice →
See all drugs & medical recalls →