Noah Medical recalls

5 official recalls and safety alerts involving Noah Medical since 2024-05, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Noah Medical or the relevant agency.

  1. 2025-12-03
  2. US
    Drugs & medicalClass II

    Noah Medical Corporation recalls Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its…

    Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.

    FDA · DeviceNoah Medical CorporationUS distribution to states of: NY, MA, WV, WY, TN, OH, NJ, MN, CA, MO, ID, IN,…Official notice →
  3. 2025-01-01
  4. US
    Drugs & medicalClass II

    Noah Medical recalls Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic…

    Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of…

    FDA · DeviceNoah MedicalUS: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NCOfficial notice →
  5. 2024-12-25
  6. US
    Drugs & medicalClass II

    Noah Medical recalls Galaxy System (GAL-001)

    Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is…

    FDA · DeviceNoah MedicalUS: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NYOfficial notice →
  7. 2024-12-18
  8. US
    Drugs & medicalClass II

    Noah Medical recalls Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)

    Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive…

    FDA · DeviceNoah MedicalUS Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ,…Official notice →
  9. 2024-05-08
  10. US
    Drugs & medicalClass II

    Noah Medical recalls Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide…

    Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.

    FDA · DeviceNoah MedicalU.S. Nationwide distribution in the states of CA, IN, MD, MO, OH, and PA.Official notice →