Noah Medical recalls Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)
Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive…
Read the official notice →- Date
- 2024-12-18
- Country
- United States
- Brand
- Noah Medical
- Product
- Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)
- Classification
- Class II
- Units
- US Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Noah Medical recalls
- 2025-12-03
US
Noah Medical Corporation recalls Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its…
Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.
FDA · DeviceNoah Medical CorporationUS distribution to states of: NY, MA, WV, WY, TN, OH, NJ, MN, CA, MO, ID, IN,…Official notice →- 2025-01-01
US
Noah Medical recalls Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic…
Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of…
FDA · DeviceNoah MedicalUS: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NCOfficial notice →- 2024-12-25
US
Noah Medical recalls Galaxy System (GAL-001)
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is…
FDA · DeviceNoah MedicalUS: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NYOfficial notice →- 2024-05-08
US
Noah Medical recalls Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide…
Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.
FDA · DeviceNoah MedicalU.S. Nationwide distribution in the states of CA, IN, MD, MO, OH, and PA.Official notice →
Other drugs & medical recalls in the United States
- 2024-12-25
US
US
Aurobindo Pharma USA Inc recalls Nebivolol Tablets, 2.5. mg, 30-count bottles, Rx only, Distributed by: Aurobindo Pharma USA, Inc., 279 Princeton-Hightstown Road, East…
CGMP Deviations: Presence of Nitrosamine Drug Substance Related Impurity (NDSRI), N-Nitroso Nebivolol above acceptable intake (AI) limit.
US
Keystone Industries recalls Gelato, Benzocaine 20% Topical Gel Anesthetic Gel, Net Wt. 1 oz. (30ml), Manufactured by Keystone Industries 480 S. Democrat Rd.,…
CGMP deviations: the bulk product was rejected by the Quality Unit after routine inspection of the mixing vessel showed scratches on the sides and bottom of the mixing vessel. Product was to be rejected but was…
FDA · DrugKeystone IndustriesNationwide in the USA and Canada, Dominican Republic, El Salvador, Honduras,…Official notice →- 2024-12-18
US
Hermes Medical Solutions AB recalls Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0
Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.
FDA · DeviceHermes Medical Solutions ABU.S.: CA, CT, FL, HI, KY, MA, MD, MO, NJ. NY, OH, TN, TX, VA, and WA O.U.S.:…Official notice →US
Diasorin Inc. recalls DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens
Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose…
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura…
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Nationwide distribution. International distribution to Albania, Algeria,…Official notice →