Noah Medical Corporation recalls Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its…
Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.
Read the official notice →- Date
- 2025-12-03
- Country
- United States
- Brand
- Noah Medical
- Product
- Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its…
- Classification
- Class II
- Units
- US distribution to states of: NY, MA, WV, WY, TN, OH, NJ, MN, CA, MO, ID, IN,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Noah Medical recalls
- 2025-01-01
US
Noah Medical recalls Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic…
Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of…
FDA · DeviceNoah MedicalUS: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NCOfficial notice →- 2024-12-25
US
Noah Medical recalls Galaxy System (GAL-001)
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is…
FDA · DeviceNoah MedicalUS: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NYOfficial notice →- 2024-12-18
US
Noah Medical recalls Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)
Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive…
FDA · DeviceNoah MedicalUS Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ,…Official notice →- 2024-05-08
US
Noah Medical recalls Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide…
Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.
FDA · DeviceNoah MedicalU.S. Nationwide distribution in the states of CA, IN, MD, MO, OH, and PA.Official notice →
Other drugs & medical recalls in the United States
- 2025-12-03
US
B Braun Medical Inc recalls Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA · DeviceB Braun Medical IncWorldwide distribution - US distribution and international distribution to…Official notice →US
BioFire Diagnostics, LLC recalls BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
FDA · DeviceBioFire Diagnostics, LLCUS:AZ, FL,GA,ID NC, NH, NJ, NY, OR, PA, SC OUS: NoneOfficial notice →US
Stryker Corporation recalls NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita;…
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
FDA · DeviceStryker CorporationWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Inpeco S.A. recalls FlexLab (FLX); Version: FLX-217-40;
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer…
FDA · DeviceInpeco S.A.US Domestic: MI. OUS International: Australia, Finland, France, Germany,…Official notice →US
Medtronic MiniMed, Inc. recalls CareLink Clinic, REF: MMT-7350
Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview…
FDA · DeviceMedtronic MiniMed, Inc.Worldwide Distribution: US (nationwide) including states of: TX, NC, MN, CA,…Official notice →US
Argon Medical Devices, Inc recalls Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the…
Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is…
FDA · DeviceArgon Medical Devices, IncU.S. (nationwide) distribution to states of: AL, AR, AZ, CA, CO, CT, DC, FL,…Official notice →