Becton, Dickinson and Company, BD Bio Sciences recalls
4 official recalls and safety alerts involving Becton, Dickinson and Company, BD Bio Sciences since 2024-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Becton, Dickinson and Company, BD Bio Sciences or the relevant agency.
- 2025-08-13
US
Becton, Dickinson and Company, BD Bio Sciences recalls BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD
Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesWorldwide - US Nationwide distribution in the states of Tennessee, New York,…Official notice →- 2025-07-09
US
Becton, Dickinson and Company, BD Bio Sciences recalls REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.
Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesUS: Arkansas, Arizona, California, Colorado, Connecticut, District of…Official notice →- 2025-01-01
US
Becton, Dickinson and Company, BD Bio Sciences recalls FACSLyric 2L4C Instrument US, Catalog Number 662875, UDI 00382906628756 BD FACSLyric" flow cytometer is intended…
On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric…
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesWorldwide - US Nationwide distribution including in the states of Alabama,…Official notice →- 2024-10-09
US
Becton, Dickinson and Company, BD Bio Sciences recalls BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System…
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated…
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesUS: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria,…Official notice →