Becton, Dickinson and Company, BD Bio Sciences recalls FACSLyric 2L4C Instrument US, Catalog Number 662875, UDI 00382906628756 BD FACSLyric" flow cytometer is intended…
On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric…
Read the official notice →- Date
- 2025-01-01
- Country
- United States
- Brand
- Becton, Dickinson and Company, BD Bio Sciences
- Product
- FACSLyric 2L4C Instrument US, Catalog Number 662875, UDI 00382906628756 BD FACSLyric" flow cytometer is intended for in vitro diagnostic…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution including in the states of Alabama,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Becton, Dickinson and Company, BD Bio Sciences recalls
- 2025-08-13
US
Becton, Dickinson and Company, BD Bio Sciences recalls BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD
Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesWorldwide - US Nationwide distribution in the states of Tennessee, New York,…Official notice →- 2025-07-09
US
Becton, Dickinson and Company, BD Bio Sciences recalls REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.
Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesUS: Arkansas, Arizona, California, Colorado, Connecticut, District of…Official notice →- 2024-10-09
US
Becton, Dickinson and Company, BD Bio Sciences recalls BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System…
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated…
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesUS: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria,…Official notice →
Other drugs & medical recalls in the United States
- 2025-01-01
US
Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates…
Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the…
FDA · DeviceAbbott MedicalWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →US
SEASPINE ORTHOPEDICS CORPORATION recalls Brand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog Number: 79-0002…
Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.
FDA · DeviceSEASPINE ORTHOPEDICS CORPORATIONU.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.Official notice →US
Straight Smile, LLC recalls BYTE Aligner System REF PRS-0063 BYTE All-Day Aligner System BYTE AT-Night Aligner System A custom-made orthodontic appliance, typically…
Virtual-only clinical workflow utilized for dental aligners and impression to approve patients for aligner devices may not be suitable for patients who have certain contraindications.
FDA · DeviceStraight Smile, LLCWorldwide - U.S. Nationwide distribution including in the states of CO, KS,…Official notice →US
Limacorporate S.p.A recalls TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.
Due to a potential difference in mechanical behavior between the implanted device and tool(s) used during revision surgery.
US
3M Company - Health Care Business recalls Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833…
Attachments may not mate with the tooth as intended in the digital treatment design.
FDA · Device3M Company - Health Care BusinessProduct was distributed nationwide US, and internationally to Canada,…Official notice →US
Noah Medical recalls Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic…
Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of…
FDA · DeviceNoah MedicalUS: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NCOfficial notice →