Becton, Dickinson and Company, BD Bio Sciences recalls REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.
Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
Read the official notice →- Date
- 2025-07-09
- Country
- United States
- Brand
- Becton, Dickinson and Company, BD Bio Sciences
- Product
- REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.
- Classification
- Class II
- Units
- US: Arkansas, Arizona, California, Colorado, Connecticut, District of…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Becton, Dickinson and Company, BD Bio Sciences recalls
- 2025-08-13
US
Becton, Dickinson and Company, BD Bio Sciences recalls BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD
Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesWorldwide - US Nationwide distribution in the states of Tennessee, New York,…Official notice →- 2025-01-01
US
Becton, Dickinson and Company, BD Bio Sciences recalls FACSLyric 2L4C Instrument US, Catalog Number 662875, UDI 00382906628756 BD FACSLyric" flow cytometer is intended…
On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric…
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesWorldwide - US Nationwide distribution including in the states of Alabama,…Official notice →- 2024-10-09
US
Becton, Dickinson and Company, BD Bio Sciences recalls BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System…
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated…
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesUS: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria,…Official notice →
Other drugs & medical recalls in the United States
- 2025-07-16
US
Breckenridge Pharmaceutical, Inc. recalls Duloxetine Delayed-Release Capsules, USP, 40mg, 30-count bottles, Rx Only, Mfr. by: Towa Pharmaceutical Europe, S.L.…
CGMP Deviations: Presence of N-nitroso-duloxetine impurity above FDA recommended interim limit
US
Macleods Pharmaceuticals Ltd recalls Famciclovir Tablets, USP 500 mg, 30-count bottles, Rx Only, Manufactured for: Macleods Pharma USA, Inc., Manufactured by: Macleods…
Presence of Foreign Substance- Black hair strand found attached to a tablet in a sealed bottle.
- 2025-07-09
US
Siemens Healthcare Diagnostics, Inc. recalls ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (ABN2 Control)
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA,…Official notice →US
Geneoscopy, Inc. recalls ColoSense Test Kit, Part No. 80-001, component of ColoSense test
Multiple plate failures documented as a result of the Low Positive Template Control being too high and out of range using the affected lot of ColoSense Test Kit.
US
C.R. Bard Inc recalls REF: 123400,Bard Urologist's Tray, For the Obstructed Urethra
Cather packaging may contain the incorrect French size.
FDA · DeviceC.R. Bard IncWorldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package…
Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →