Zyno Medical LLC recalls Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having…
Read the official notice →- Date
- 2024-07-10
- Country
- United States
- Brand
- Zyno Medical
- Product
- Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Zyno Medical recalls
- 2025-06-18
US
Zyno Medical LLC recalls Z-800 Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Unreleased software versions were installed on distributed devices without verification or validation.
- 2024-12-04
US
Zyno Medical LLC recalls Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide…
For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
- 2024-11-27
US
Zyno Medical LLC recalls Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous…
Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
- 2024-10-16
US
Zyno Medical LLC recalls Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and…
There is a defect in the air-in-line software algorithm.
- 2024-03-06
US
Zyno Medical LLC recalls Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to…
When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in…
Other drugs & medical recalls in the United States
- 2024-07-10
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS…
VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of FL, IL, NY, PA and the…Official notice →US
Abbott Medical recalls Abbott Liberta RC, Implantable Pulse Generator, REF: 62400, STERILE, RxONLY
Deep brain stimulation system will first turn off after approximately 50 days after the system is activated and subsequently, this same action will recur approximately every 50 days thereafter. This unexpected lost of…
FDA · DeviceAbbott MedicalUS: Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GA,…Official notice →US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic
Beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found…
US
CareFusion 303, Inc. recalls BD Alaris System Manager REF: 9601, compatible with BD Alaris PC Units
Due to a software issue the PC unit may not connect to the server which could impact wireless data transmission to and from the server.
FDA · DeviceCareFusion 303, Inc.US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT…Official notice →US
Stryker Corporation recalls The Non-Sterile Flyte hood covers the user s head and neck region and is intended to be worn with commercially available gowns. The hood…
Expired Products distributed to customers
FDA · DeviceStryker CorporationUS Nationwide distribution in the states of Arkansas, California, Colorado,…Official notice →US
GE Medical Systems Information Technologies Inc recalls Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab…
Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.
FDA · DeviceGE Medical Systems Information Technologies IncWorldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL,…Official notice →