Drugs & medicalClass I

Zyno Medical LLC recalls Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and…

There is a defect in the air-in-line software algorithm.

Read the official notice →
Date
2024-10-16
Country
United States
Brand
Zyno Medical
Product
Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and…
Classification
Class I
Units
US Nationwide.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Zyno Medical recalls

  1. 2025-06-18
  2. US
    Drugs & medicalClass I

    Zyno Medical LLC recalls Z-800 Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

    Unreleased software versions were installed on distributed devices without verification or validation.

    FDA · DeviceZyno Medical LLCUS NationwideOfficial notice →
  3. 2024-12-04
  4. US
    Drugs & medicalClass III

    Zyno Medical LLC recalls Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide…

    For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.

    FDA · DeviceZyno Medical LLCDomestic: AL, FL, LA, MI, NE, NJ, OH, TX.Official notice →
  5. 2024-11-27
  6. US
    Drugs & medicalClass II

    Zyno Medical LLC recalls Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous…

    Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.

    FDA · DeviceZyno Medical LLCThis product is distributed only in the United States.Official notice →
  7. 2024-07-10
  8. US
    Drugs & medicalClass II

    Zyno Medical LLC recalls Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps

    Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having…

    FDA · DeviceZyno Medical LLCUS Nationwide distribution.Official notice →
  9. 2024-03-06
  10. US
    Drugs & medicalClass II

    Zyno Medical LLC recalls Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to…

    When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in…

    FDA · DeviceZyno Medical LLCNationwideOfficial notice →
All 6 Zyno Medical recalls →

Other drugs & medical recalls in the United States

  1. 2024-10-16
  2. US
    Drugs & medicalClass II

    Qualigen Inc recalls FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

    Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

    FDA · DeviceQualigen IncUS: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT,…Official notice →
  3. US
    Drugs & medicalClass II

    Stryker Neurovascular recalls The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end…

    Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed…

    FDA · DeviceStryker NeurovascularU.S. Nationwide distribution in the states of CA and VA.Official notice →
  4. US
    Drugs & medicalClass II

    Philips North America Llc recalls IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 are affected as follows:…

    Devices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices.…

    FDA · DevicePhilips North America LlcForeign Only: Denmark France Germany Japan United KingdomOfficial notice →
  5. US
    Drugs & medicalClass I

    123Herbals recalls Vail-Bon Jie Yang Wan capsules, 30-count bottle, Vall Boon Medical Factory Sdn. Bhd., No 1, Lorong Mengkudu, Taman Mengkudu, 14100 Juru,…

    Marketed without an approved NDA/ANDA: Product found to be tainted with undeclared dexamethasone and chlorpheniramine.

    FDA · Drug123HerbalsDistributed via Amazon and 123Herbals.com websiteOfficial notice →
  6. 2024-10-09
  7. US
    Drugs & medicalClass II

    Alcon Research LLC recalls DAILIES TOTAL1, DAILIES TOTAL1 for Astigmatism, DAILIES TOTAL1 Multifocal, TOTAL30 Sphere,

    Material with a quality issue was used in contact lens production, so contact lenses may not meet quality and/or performance standards for the entirety of their labeled shelf life, and wearing these lenses may lead to…

    FDA · DeviceAlcon Research LLCUS: UT, TX, NC, MT, AL, WA, MS, PA, LA, MI, GA, FL, KY, OH, AZ, IL, CA, NJ,…Official notice →
  8. US
    Drugs & medicalClass II

    Philips North America Llc recalls DIVA 24 Inch Widescreen LCD Touch Display - Used with the Philips PIC iX system central monitoring system. Model: RGD2461AMI Philips…

    If the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a…

    FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution in the states of CA, IL, ME, NC, NJ,…Official notice →