Zyno Medical LLC recalls Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to…
When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in…
Read the official notice →- Date
- 2024-03-06
- Country
- United States
- Brand
- Zyno Medical
- Product
- Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to…
- Classification
- Class II
- Units
- Nationwide
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Zyno Medical recalls
- 2025-06-18
US
Zyno Medical LLC recalls Z-800 Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Unreleased software versions were installed on distributed devices without verification or validation.
- 2024-12-04
US
Zyno Medical LLC recalls Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide…
For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
- 2024-11-27
US
Zyno Medical LLC recalls Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous…
Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
- 2024-10-16
US
Zyno Medical LLC recalls Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and…
There is a defect in the air-in-line software algorithm.
- 2024-07-10
US
Zyno Medical LLC recalls Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having…
Other drugs & medical recalls in the United States
- 2024-03-06
US
Abbott Molecular, Inc. recalls Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)
Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.
FDA · DeviceAbbott Molecular, Inc.US Nationwide distribution in the states of AL, CA, IL, MA, MD, MN, NY, and PA.Official notice →US
EMD Millipore Corporation recalls Schiff Reagent, Hotchkiss-McManus, packaged in an amber glass bottle with a black polypropylene cap.
This IVD is colorless. A customer complaint was received noting it appeared pink. A review of inventory confirmed additional stock appeared pink.
FDA · DeviceEMD Millipore CorporationDomestic: AL, IL, MA, MN, NJ, PA, TX, VA. Foreign: Canada.Official notice →US
Medagent Inc recalls Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 06-06-285 and 06-06-285MOD.…
Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of patient's safety if the movement is transferred to the spatula, which is attached to the holding arm.
FDA · DeviceMedagent IncTX Foreign: Algeria China Dominican Republic Germany India Japan Philippines UKOfficial notice →US
Alphatec Spine, Inc. recalls IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 132-10-85-300…
Due to reports of intraoperative graft bolt implantation breakages.
FDA · DeviceAlphatec Spine, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AZ,…Official notice →US
Medtronic Neurosurgery recalls Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, REF 46913, Sterile EO, Rx Only. Used to drain…
Potential for catheter disconnection from the patient line stopcock connectors.
FDA · DeviceMedtronic NeurosurgeryWorldwide Distribution: US (nationwide): AL, AR, AZ, CA, CO, CT, DC, DE, FL,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE TRAY TRACH BASIC SOLUTION, REF DYND40511
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline…
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldDistribution US nationwide including Puerto Rico, Canada, UAE, and PanamaOfficial notice →