Drugs & medicalClass II

Spacelabs Healthcare, Inc. recalls Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The…

Due to off-the-shelf display port component that has audio/video failures

Read the official notice →
Date
2026-08-19
Country
United States
Brand
Spacelabs Healthcare
Product
Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Spacelabs Healthcare recalls

  1. 2026-05-13
  2. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A…

    Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide - US Nationwide di GA, KS, MI, OH, SC, TX, WA and the country of…Official notice →
  3. 2026-04-15
  4. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

    Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO…

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide - US Nationwide distribution including in the states of NY, NJ, AL,…Official notice →
  5. 2025-11-12
  6. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Ltd. recalls Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description:…

    Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific…

    FDA · DeviceSpacelabs Healthcare, Ltd.Worldwide - US Nationwide distribution in the state of MN and the countries of…Official notice →
  7. 2025-10-22
  8. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software…

    Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of…

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide - United States Nationwide distribution including in the states of…Official notice →
  9. 2025-01-15
  10. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.

    Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…

    FDA · DeviceSpacelabs Healthcare, Inc.US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL,…Official notice →
  11. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

    Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…

    FDA · DeviceSpacelabs Healthcare, Inc.US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX,…Official notice →
All 10 Spacelabs Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2026-08-19
  2. US
    Drugs & medicalClass II

    Medline Industries, LP recalls Medline Micro-Kill Bleach Germicidal Bleach Wipes, 12" x 12", 110 Count Bag, Product Number MSC351450BG4

    Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may…

    FDA · DeviceMedline Industries, LPUS Nationwide distribution in the states and territories of Guam, Puerto Rico.Official notice →
  3. US
    Drugs & medicalClass II

    Covidien, LP recalls V-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1 BOX BYPASS CARDON…

    Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or…

    FDA · DeviceCovidien, LPUS Nationwide and the countries of Australia, Austria, Belgium, Bulgaria,…Official notice →
  4. US
    Drugs & medicalClass II

    Galt Medical Corporation recalls Boston Scientific Enhance Transcarotid/Peripheral Access Kit REF: SR-4F21G7D-MP (Kit-075-03) Sheath with Dilator , Stiffen Dilator,…

    Black depth markings may detach from the outer sheath of the coaxial dilator introducer component of transcarotid-peripheral access kits.

    FDA · DeviceGalt Medical CorporationUS Nationwide distribution in the state of MN.Official notice →
  5. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1010. Flexible, self-expanding endoprosthesis designed to maintain…

    10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).

    FDA · DeviceW L Gore & Associates, Inc.U.S. Nationwide distribution in the states of GA and RI.Official notice →
  6. US
    Drugs & medicalClass II

    Ventana Medical Systems, Inc. recalls navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number…

    Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window

    FDA · DeviceVentana Medical Systems, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  7. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls BOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in facilitating the insertion and…

    Due to an increase rate of balloon pinhole defects.

    FDA · DeviceMICROVENTION INC.Worldwide - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →
See all drugs & medical recalls →