Drugs & medicalClass II

Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…

Read the official notice →
Date
2025-01-15
Country
United States
Brand
Spacelabs Healthcare
Product
Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.
Classification
Class II
Units
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Spacelabs Healthcare recalls

  1. 2025-11-12
  2. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Ltd. recalls Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description:…

    Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific…

    FDA · DeviceSpacelabs Healthcare, Ltd.Worldwide - US Nationwide distribution in the state of MN and the countries of…Official notice →
  3. 2025-10-22
  4. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software…

    Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of…

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide - United States Nationwide distribution including in the states of…Official notice →
  5. 2025-01-15
  6. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.

    Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…

    FDA · DeviceSpacelabs Healthcare, Inc.US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL,…Official notice →
  7. 2024-12-25
  8. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Ltd. recalls Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory…

    Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the…

    FDA · DeviceSpacelabs Healthcare, Ltd.U.S.: AR, CA, CO, DE, FL, FA, LA, MA, MI, MN, MT, NC, NE, NU OH, OR, PA,…Official notice →
  9. 2023-05-10
  10. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Ultraview SL (UVSL) Command Module, Model 91496

    Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent…

    FDA · DeviceSpacelabs Healthcare, Inc.US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR,…Official notice →
  11. 2022-08-17
  12. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver, Model: 96280

    Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal…

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide distribution - US Nationwide distribution in the states of NY, AR,…Official notice →
All 10 Spacelabs Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2025-01-15
  2. US
    Drugs & medicalClass II

    GE Medical Systems, LLC recalls Tec 820 ISO Anesthesia Vaporizer, Reference Number 1177-9820-ISO

    Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.

    FDA · DeviceGE Medical Systems, LLCUS Nationwide. Global Distribution.Official notice →
  3. US
    Drugs & medicalClass II

    GE Medical Systems, LLC recalls GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB;…

    GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and…

    FDA · DeviceGE Medical Systems, LLCWorldwideOfficial notice →
  4. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.

    Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…

    FDA · DeviceSpacelabs Healthcare, Inc.US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL,…Official notice →
  5. US
    Drugs & medicalClass II

    Aerin Medical, Inc. recalls Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.

    Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.

    FDA · DeviceAerin Medical, Inc.US Nationwide Distribution: AL, AZ, CA, CO, GA, ID, IL, IN, KY LA, MI, MN, NY,…Official notice →
  6. US
    Drugs & medicalClass II

    Stryker Corporation recalls 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V…

    Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of…

    FDA · DeviceStryker CorporationUS (Domestic) distribution to states of: WA, GA, and IL. OUS (International)…Official notice →
  7. US
    Drugs & medicalClass II

    Elucent Medical Inc recalls EnVisio Patient Pad Transition Cushion, Model/Catalog Number: ENVISIO-SYS Product Description: The EnVisio Patient Pad Transition Cushion…

    There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation…

    FDA · DeviceElucent Medical IncUS Distribution: AZ, CA, CO, CT, FL, GA, IL, KY, MN, NC, ND, NJ, NY, OH, PA,…Official notice →
See all drugs & medical recalls →