Drugs & medicalClass II

Spacelabs Healthcare, Inc. recalls Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A…

Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).

Read the official notice →
Date
2026-05-13
Country
United States
Brand
Spacelabs Healthcare
Product
Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A…
Classification
Class II
Units
Worldwide - US Nationwide di GA, KS, MI, OH, SC, TX, WA and the country of…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Spacelabs Healthcare recalls

  1. 2026-08-19
  2. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The…

    Due to off-the-shelf display port component that has audio/video failures

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA,…Official notice →
  3. 2026-04-15
  4. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

    Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO…

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide - US Nationwide distribution including in the states of NY, NJ, AL,…Official notice →
  5. 2025-11-12
  6. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Ltd. recalls Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description:…

    Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific…

    FDA · DeviceSpacelabs Healthcare, Ltd.Worldwide - US Nationwide distribution in the state of MN and the countries of…Official notice →
  7. 2025-10-22
  8. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software…

    Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of…

    FDA · DeviceSpacelabs Healthcare, Inc.Worldwide - United States Nationwide distribution including in the states of…Official notice →
  9. 2025-01-15
  10. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.

    Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…

    FDA · DeviceSpacelabs Healthcare, Inc.US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL,…Official notice →
  11. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

    Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…

    FDA · DeviceSpacelabs Healthcare, Inc.US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX,…Official notice →
All 10 Spacelabs Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2026-05-13
  2. US
    Drugs & medicalClass II

    BioFire Diagnostics, LLC recalls BioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.

    Contamination to in-vitro diagnostic test may result in false positives.

    FDA · DeviceBioFire Diagnostics, LLCWorldwide - US Nationwide distribution in the states of TX, NC, UT, MD and the…Official notice →
  3. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957,…

    Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's…

    FDA · DevicePhysio-Control, Inc.US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA,…Official notice →
  4. US
    Drugs & medicalClass II

    Erbe Medical, LLC recalls ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus…

    Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported…

    FDA · DeviceErbe Medical, LLCUS Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY,…Official notice →
  5. US
    Drugs & medicalClass II

    Centinel Spine, Inc. recalls Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.

    Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.

    FDA · DeviceCentinel Spine, Inc.US Nationwide distribution in the states of AZ, CA, GA, LA, MO, NY, TN, TX.Official notice →
  6. US
    Drugs & medicalClass II

    Boston Scientific Corporation recalls LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5

    For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In…

    FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide and the country of Belgium.Official notice →
  7. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software…

    Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.

    FDA · DeviceIntuitive Surgical, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
See all drugs & medical recalls →