Siemens Medical Solutions USA, Inc recalls Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641…
System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances,…
Read the official notice →- Date
- 2021-11-24
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641 (3)Sensis Vibe…
- Classification
- Class II
- Units
- Nationwide
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2022-01-26
US
Siemens Medical Solutions USA, Inc recalls 1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis…
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI,…Official notice →- 2022-01-19
US
Siemens Medical Solutions USA, Inc recalls SOMATOM Edge Plus-Computed tomography system Model 1026700
Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.
- 2022-01-05
US
Siemens Medical Solutions USA, Inc recalls ARTISTE with syngo RT Therapist-linear accelerator systems to deliver X-Ray photon and electron radiation for the therapeutic…
Potential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of IA, LA, NE, NJ, NV, NY, OH, PA,…Official notice →- 2021-12-01
US
Siemens Medical Solutions USA, Inc recalls Artis Q.zen ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and…
If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
US
Siemens Medical Solutions USA, Inc recalls SOMATOM Definition Edge with software syngo.CT VB20 Model #10590000
software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient…
- 2021-11-10
US
Siemens Medical Solutions USA, Inc recalls Artis zeego, Fluoroscopic X-Ray System Models: 10280959,10502505, 10848283 in combination with the following workplace/system…
Software error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown…
Other drugs & medical recalls in the United States
- 2021-11-24
US
Bio-Rad Laboratories, Inc. recalls BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is…
Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation…
FDA · DeviceBio-Rad Laboratories, Inc.Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA,…Official notice →US
Arrow International Inc recalls Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-W The Arrow-Trerotola" Percutaneous…
7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.
FDA · DeviceArrow International IncWorldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: BE…Official notice →US
Beaver Visitec International, Inc. recalls bvi CustomEyes Kit-Lafayette General Medical Center: CustomEyes Cataract Surgical Kits that contain Beaver Xstar Crescent…
Product packaged as Beaver Xstar Crescent Knife, 2.5mm, 55 degrees bevel-down blades contain bevel-up blades
FDA · DeviceBeaver Visitec International, Inc.Nationwide Foreign: Australia Austria France Germany Netherlands Spain Japan…Official notice →US
Scrip Inc recalls Scrip Ultrasound Gel Clear, 5L Intended for use in acoustic coupling of ultrasound transducer to tissue
Ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination.
US
Cook Inc. recalls Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal…
Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time…
FDA · DeviceCook Inc.Domestic distribution nationwide. Foreign distribution worldwide.Official notice →US
Cordis Corporation recalls SMART FLEX 6x150 BIL, 120cm The S.M.A.R.T.¿ Flex Biliary Stent System is indicated for use in the palliation of malignant strictures in…
There is a potential for distal tip dislodgement or separation.
FDA · DeviceCordis CorporationUS Distribution to states of: CA, FL, IN, LA, OH, OK, TX and OUS distribution…Official notice →