Siemens Medical Solutions USA, Inc recalls Artis zeego, Fluoroscopic X-Ray System Models: 10280959,10502505, 10848283 in combination with the following workplace/system…
Software error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown…
Read the official notice →- Date
- 2021-11-10
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Artis zeego, Fluoroscopic X-Ray System Models: 10280959,10502505, 10848283 in combination with the following workplace/system application…
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2022-01-26
US
Siemens Medical Solutions USA, Inc recalls 1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis…
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI,…Official notice →- 2022-01-19
US
Siemens Medical Solutions USA, Inc recalls SOMATOM Edge Plus-Computed tomography system Model 1026700
Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.
- 2022-01-05
US
Siemens Medical Solutions USA, Inc recalls ARTISTE with syngo RT Therapist-linear accelerator systems to deliver X-Ray photon and electron radiation for the therapeutic…
Potential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of IA, LA, NE, NJ, NV, NY, OH, PA,…Official notice →- 2021-12-01
US
Siemens Medical Solutions USA, Inc recalls Artis Q.zen ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and…
If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
US
Siemens Medical Solutions USA, Inc recalls SOMATOM Definition Edge with software syngo.CT VB20 Model #10590000
software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient…
- 2021-11-24
US
Siemens Medical Solutions USA, Inc recalls Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641…
System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances,…
Other drugs & medical recalls in the United States
- 2021-11-10
US
Jjgc Industria E Comercio De Materials Dentarios Sa recalls GM Helix Acqua Implant 3.5x10, Article No. 140.944
3.5x11.5mm implants were mislabeled as 3.5x10mm implants. In borderline cases where the bone quality is low and the bone ridge has a maximum indication for the implant height (10mm), if the incorrect implant is…
FDA · DeviceJjgc Industria E Comercio De Materials Dentarios SaUS Nationwide distribution.Official notice →US
MICROVENTION INC. recalls Aspiration Syringe Kit, REF: MVSK60
Syringe Kits labeled for distribution OUS were inadvertently shipped to US customers.
FDA · DeviceMICROVENTION INC.US Nationwide distribution in the states of AL, NY, NV.Official notice →US
Medtronic Neuromodulation recalls Clinician Programmer Application (CPA) model A610 Clinician Software Application
A software anomaly may occur with the clinician programmer application.
FDA · DeviceMedtronic NeuromodulationWorldwide distribution - US Nationwide distribution and the countries of…Official notice →US
Bayer Medical Care, Inc. recalls The MEDRAD Twist & Go Disposable Syringe 150 mL with Quick Fill Tube is a syringe made from a clear polycarbonate material and is an…
Labeling error: MEDRAD Mark 7 Arterion Injection System Syringes (ART 700 SYR) labeled as Twist & Go Syringes (TAG 150 SYR)
FDA · DeviceBayer Medical Care, Inc.Worldwide distribution - US Nationwide and the countries of AU, BY, DE, DK,…Official notice →US
Implant Direct Sybron Manufacturing LLC recalls ImplantDirect, REF 834713, Legacy 2 Implant 4.7mmD x 13mmL, 4.5mmD Platform,
Packaged dental implant contains a different size then the size declared on the labeling.
FDA · DeviceImplant Direct Sybron Manufacturing LLCWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →US
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur SARS-CoV-2 Antigen Assay (100 Test Kit), SMN # 11207866
Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine that the assays were adequately validated for the intended use.
FDA · DeviceSiemens Healthcare Diagnostics, IncUS Nationwide distribution in the states of AK, CA, CO, DC, GA, IL, ME, NC,…Official notice →