Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur SARS-CoV-2 Antigen Assay (100 Test Kit), SMN # 11207866

Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine that the assays were adequately validated for the intended use.

Read the official notice →
Date
2021-11-10
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
ADVIA Centaur SARS-CoV-2 Antigen Assay (100 Test Kit), SMN # 11207866
Classification
Class II
Units
US Nationwide distribution in the states of AK, CA, CO, DC, GA, IL, ME, NC,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2022-01-12
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH 930 Analyzer with Software version: V1.25.1 and lower- An automated, clinical chemistry analyzer designed to…

    (1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: AU DE AE TR BG CL IN CZ BR ES KE CN PL SA BE IQ IT CW SE…Official notice →
  3. 2022-01-05
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of…

    ¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, CO,…Official notice →
  5. 2021-12-08
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 500- In-vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting,…

    Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution and the countries of DE,…Official notice →
  7. 2021-11-10
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1…

    The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution.Official notice →
  9. 2021-11-03
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151

    The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution.Official notice →
  11. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11097543

    The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution.Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2021-11-10
  2. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Artis zeego, Fluoroscopic X-Ray System Models: 10280959,10502505, 10848283 in combination with the following workplace/system…

    Software error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Bayer Medical Care, Inc. recalls The MEDRAD Twist & Go Disposable Syringe 150 mL with Quick Fill Tube is a syringe made from a clear polycarbonate material and is an…

    Labeling error: MEDRAD Mark 7 Arterion Injection System Syringes (ART 700 SYR) labeled as Twist & Go Syringes (TAG 150 SYR)

    FDA · DeviceBayer Medical Care, Inc.Worldwide distribution - US Nationwide and the countries of AU, BY, DE, DK,…Official notice →
  4. US
    Drugs & medicalClass II

    Implant Direct Sybron Manufacturing LLC recalls ImplantDirect, REF 834713, Legacy 2 Implant 4.7mmD x 13mmL, 4.5mmD Platform,

    Packaged dental implant contains a different size then the size declared on the labeling.

    FDA · DeviceImplant Direct Sybron Manufacturing LLCWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
  5. US
    Drugs & medicalClass I

    Lumenis, Inc. recalls IPL Coupling Gel, I L (AX1009013), 12 pack of 0.25 L (AX1009018), 6 pack of 1 L (AX1009019 or AX1009063); and IPL Coupling Gel or EcoGel…

    Ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021, FDA published a Letter to Health Care Providers to further address this issue:…

    FDA · DeviceLumenis, Inc.Worldwide distribution - US Nationwide distribution in the states of KS, PR,…Official notice →
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1…

    The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    FujiFilm Healthcare Americas Corporation recalls Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0

    When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow…

    FDA · DeviceFujiFilm Healthcare Americas CorporationUS Nationwide distribution.Official notice →
See all drugs & medical recalls →