Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of…

¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on…

Read the official notice →
Date
2022-01-05
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human…
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the states of CA, CO,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2022-04-20
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Fructosamine (Fruc), Siemens Material Number 11097637

    Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Global Distribution. US Nationwide.Official notice →
  3. 2022-04-06
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension…

    Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution: US (nationwide) 47 States including: AK, AL, AR, AZ,…Official notice →
  5. 2022-01-12
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH 930 Analyzer with Software version: V1.25.1 and lower- An automated, clinical chemistry analyzer designed to…

    (1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: AU DE AE TR BG CL IN CZ BR ES KE CN PL SA BE IQ IT CW SE…Official notice →
  7. 2021-12-08
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 500- In-vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting,…

    Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution and the countries of DE,…Official notice →
  9. 2021-11-10
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur SARS-CoV-2 Antigen Assay (100 Test Kit), SMN # 11207866

    Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine that the assays were adequately validated for the intended use.

    FDA · DeviceSiemens Healthcare Diagnostics, IncUS Nationwide distribution in the states of AK, CA, CO, DC, GA, IL, ME, NC,…Official notice →
  11. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1…

    The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution.Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2022-01-05
  2. US
    Drugs & medicalClass II

    AMO Manufacturing USA, LLC recalls iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive…

    Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations.…

    FDA · DeviceAMO Manufacturing USA, LLCUS: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY,…Official notice →
  3. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA TRAY (STERRAD, INLINE CAMERA…

    To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and…

    FDA · DeviceKarl Storz EndoscopyUS: AK, AL, AP, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS,…Official notice →
  4. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTISTE with syngo RT Therapist-linear accelerator systems to deliver X-Ray photon and electron radiation for the therapeutic…

    Potential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of IA, LA, NE, NJ, NV, NY, OH, PA,…Official notice →
  5. US
    Drugs & medicalClass II

    bioMerieux, Inc. recalls VIDAS RUB IgG (RBG), REF 30226

    bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.

    FDA · DevicebioMerieux, Inc.US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Forbes Rehab Services Inc recalls Affected components are limited to the Monoprice charging cable (Product # 4867) and 24W iClever Boost Cube charging brick components…

    Warning Added: Using damaged cables or chargers, or charging when moisture is present, can cause fire, electric shock, injury, or damage to Speech Generating Device or other property

    FDA · DeviceForbes Rehab Services IncUS Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Biomerieux Inc recalls RAPID 20 E, REF 20701, System for the identification of Enterobacteriaceae in 4 hours, 25 strips/package, IVD.

    There is an error in the new version of the package insert. The Reading Table lists the incorrect color for a negative result for the CIT test.

    FDA · DeviceBiomerieux IncUS Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA,…Official notice →
See all drugs & medical recalls →