Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151

The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c…

Read the official notice →
Date
2021-11-03
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2022-01-05
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of…

    ¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, CO,…Official notice →
  3. 2021-12-08
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 500- In-vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting,…

    Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution and the countries of DE,…Official notice →
  5. 2021-11-10
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur SARS-CoV-2 Antigen Assay (100 Test Kit), SMN # 11207866

    Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine that the assays were adequately validated for the intended use.

    FDA · DeviceSiemens Healthcare Diagnostics, IncUS Nationwide distribution in the states of AK, CA, CO, DC, GA, IL, ME, NC,…Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1…

    The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution.Official notice →
  8. 2021-11-03
  9. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11097543

    The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution.Official notice →
  10. 2021-10-27
  11. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc recalls Atellica IM 1600 Analyzer, SMN 11066000

    Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result,…

    FDA · DeviceSiemens Healthcare Diagnostics, IncDomestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2021-11-10
  2. US
    Drugs & medicalClass II

    Dr. Reddy's Laboratories, Inc. recalls Dr. Reddy's Ezetimibe and Simvastatin Tablets, 10 mg/40 mg, 500 count bottles, Rx Only, Manufactured by: Dr. Reddy's Laboratories…

    Failed Excipient Specifications and Presence of Foreign Tablets/Capsules; product manufactured using an excipient found to be OOS for conductivity and some Ezetimibe and Simvastatin Tablets, 10 mg/10 mg were found in…

    FDA · DrugDr. Reddy's Laboratories, Inc.Nationwide in the USOfficial notice →
  3. US
    Drugs & medicalClass II

    Dr. Reddy's Laboratories, Inc. recalls Dr. Reddy's Ezetimibe and Simvastatin Tablets, 10 mg/40 mg, a) 30 count (NDC 43598-743-30), b) 90 count (NDC 43598-743-90)…

    Failed Excipient Specifications; product manufactured using an excipient found to be OOS for conductivity

    FDA · DrugDr. Reddy's Laboratories, Inc.Nationwide in the USOfficial notice →
  4. 2021-11-03
  5. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00

    Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.

    FDA · DeviceElekta, Inc.Distribution throughout USA OUS distribution including:Algeria, Argentina,…Official notice →
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11097543

    The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Syringe Tip Cap Bulk Sterile Convenience Pak-1000 units/shelf carton. Used to seal syringes and maintain container closure…

    Holes in product packaging impacts package integrity and may compromise sterility.

    FDA · DeviceBecton Dickinson & CompanyTXOfficial notice →
  8. US
    Drugs & medicalClass II

    Smiths Medical ASD Inc. recalls Bivona Aire-Cuf Endotracheal Tube

    Inadequate pouch seal leading to a compromised sterile barrier on products with a shaft length of 325mm.

    FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution - US Nationwide distribution in the states of AL, AZ,…Official notice →
See all drugs & medical recalls →