Bio-Rad Laboratories, Inc. recalls BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is…
Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation…
Read the official notice →- Date
- 2021-11-24
- Country
- United States
- Brand
- Bio-Rad Laboratories
- Product
- BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended…
- Classification
- Class II
- Units
- Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Bio-Rad Laboratories recalls
- 2023-04-19
US
Bio-Rad Laboratories, Inc. recalls BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots.…
FDA · DeviceBio-Rad Laboratories, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Czech…Official notice →- 2022-12-07
US
Bio-Rad Laboratories, Inc. recalls CMV IgM EIA, in vitro diagnostic.
Due to an unusual increase in the positivity rate with human IgM antibodies to cytomegalovirus (CMV) Enzyme Immunoassay (EIA).
FDA · DeviceBio-Rad Laboratories, Inc.U.S.: CA, CO, FL, GA, HI, IA, MA, MN, NC, NH, NJ, PA, TN, and TX O.U.S.: NoneOfficial notice →- 2022-06-08
US
Bio-Rad Laboratories, Inc. recalls Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000
The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in…
FDA · DeviceBio-Rad Laboratories, Inc.US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH,…Official notice →- 2022-05-25
US
Bio-Rad Laboratories, Inc. recalls BioPlex 2200 REF 12000670 ToRC IgM Reagent Pack
Performance concerns for the CMV IgM analyte. Analyte has demonstrated variable elevated background depending on the handling of the reagent pack.
FDA · DeviceBio-Rad Laboratories, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AR, AZ,…Official notice →- 2022-03-23
US
Bio-Rad Laboratories, Inc. recalls BioPlex 2200 REF 12000650 SYPH T PACK Syphilis Total & RPR Reagent Pack
Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that…
FDA · DeviceBio-Rad Laboratories, Inc.Worldwide distribution U.S. Nationwide distribution in the states of AK, AL,…Official notice →- 2021-10-20
US
Bio-Rad Laboratories, Inc. recalls BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; and ANA Screen…
A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.
FDA · DeviceBio-Rad Laboratories, Inc.Worldwide distribution - U.S Nationwide distribution including in the states…Official notice →
Other drugs & medical recalls in the United States
- 2021-11-24
US
Laboratoires Bodycad, Inc. recalls Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable…
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
FDA · DeviceLaboratoires Bodycad, Inc.Distribution was made to CO, MI, and NY. Foreign distribution was made to…Official notice →US
bioMerieux, Inc. recalls API 50 CH, Model 50300
There is potential to provide an incorrect organism identification.
US
OmniLife Science recalls OMNI K2 Hip Stem, Manufacturer's Product Code 130006 The OMNI K2 Hip Stem is intended for use as the femoral component of a primary total…
The device is missing grit blasting on the proximal surface of the femoral stem, which may lead to the possibility of product revision in the event of undersizing.
US
Arrow International Inc recalls Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-W The Arrow-Trerotola" Percutaneous…
7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.
FDA · DeviceArrow International IncWorldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: BE…Official notice →US
Beaver Visitec International, Inc. recalls bvi CustomEyes Kit-Lafayette General Medical Center: CustomEyes Cataract Surgical Kits that contain Beaver Xstar Crescent…
Product packaged as Beaver Xstar Crescent Knife, 2.5mm, 55 degrees bevel-down blades contain bevel-up blades
FDA · DeviceBeaver Visitec International, Inc.Nationwide Foreign: Australia Austria France Germany Netherlands Spain Japan…Official notice →US
Siemens Medical Solutions USA, Inc recalls Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641…
System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances,…