Siemens Medical Solutions USA, Inc. recalls Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system…
The clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study…
Read the official notice →- Date
- 2022-03-23
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system intended for the…
- Classification
- Class II
- Units
- Worldwide distribution: US (nationwide) and countries of: Algeria, Andorra,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2022-04-13
US
Siemens Medical Solutions USA, Inc. recalls Symbia Intevo Bold and Symbia Intevo 6 SPECT/CT Systems
There is a quality issue with the three bolts that hold the CT balancing weights, which could create a safety issue.
FDA · DeviceSiemens Medical Solutions USA, Inc.Distribution OUS only to El Salvador, France, Germany, Japan, Taiwan, and UK.Official notice →US
Siemens Medical Solutions USA, Inc recalls YSIO X.Pree
For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide, Worldwide distributionOfficial notice →- 2022-04-06
US
Siemens Medical Solutions USA, Inc recalls Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system
Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If…
- 2022-02-09
US
Siemens Medical Solutions USA, Inc. recalls ACUSON Sequoia Ultrasound Imaging System with Software Version VA25A/B
Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →- 2022-02-02
US
Siemens Medical Solutions USA, Inc recalls Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000
Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be…
- 2022-01-26
US
Siemens Medical Solutions USA, Inc recalls 1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis…
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI,…Official notice →
Other drugs & medical recalls in the United States
- 2022-03-23
US
Boston Scientific Corporation recalls HYDRATOME RX 44-20MM/450CM Material Number: M00583060
Sterility of device is compromised due to a sterile barrier breach
FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Hologic, Inc recalls Panther Fusion Extraction Reagent-X, REF PRD-04477, For use with Panther Fusion System, IVD, Rx only, Contents: FCR-X Fusion Capture…
Failed stability specifications for reagent kits that may result in false negative test results.
FDA · DeviceHologic, IncWorldwide distribution - US Nationwide distribution in the states of AR, CA,…Official notice →US
Abbott Laboratories, Inc recalls Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro…
Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.
FDA · DeviceAbbott Laboratories, IncWorldwide distribution: US (nationwide) including states of: AL, CA, GA, IL,…Official notice →US
SD Biosensor, Inc. recalls STANDARD Q COVID-19 Ag Home Test STANDARD Q COVID-19 Ag Home Test is a rapid chromatographic immunoassay for the qualitative detection of…
Illegal importation of COVID-19 AG Test kits into the U.S. The COVID-19 Ag Test are not approved, cleared or authorized for distribution within the U.S.
US
Walmart Stores recalls Avaphinal Maximum Male Enhancement Sexual Pills, 2000 MG*, packaged in 10 Capsules per carton.
Marketed Without An Approved NDA/ANDA: Product was found to contain undeclared sildenafil, an ingredient found in FDA approved products for the treatment of male sexual enhancement, making this an unapproved drug.
US