Siemens Medical Solutions USA, Inc recalls Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system
Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If…
Read the official notice →- Date
- 2022-04-06
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system
- Classification
- Class II
- Units
- US Nationwide.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2022-04-27
US
Siemens Medical Solutions USA, Inc recalls SIEMENS SOMATOM with software syngo CT VA30A_SP4 Models: SOMATOM go.Now 11061610 & 11061618 SOMATOM go.Up 11061620 & 11061628…
Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows.…
- 2022-04-13
US
Siemens Medical Solutions USA, Inc. recalls Symbia Intevo Bold and Symbia Intevo 6 SPECT/CT Systems
There is a quality issue with the three bolts that hold the CT balancing weights, which could create a safety issue.
FDA · DeviceSiemens Medical Solutions USA, Inc.Distribution OUS only to El Salvador, France, Germany, Japan, Taiwan, and UK.Official notice →US
Siemens Medical Solutions USA, Inc recalls YSIO X.Pree
For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide, Worldwide distributionOfficial notice →- 2022-03-23
US
Siemens Medical Solutions USA, Inc. recalls Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system…
The clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study…
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution: US (nationwide) and countries of: Algeria, Andorra,…Official notice →- 2022-02-09
US
Siemens Medical Solutions USA, Inc. recalls ACUSON Sequoia Ultrasound Imaging System with Software Version VA25A/B
Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →- 2022-02-02
US
Siemens Medical Solutions USA, Inc recalls Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000
Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be…
Other drugs & medical recalls in the United States
- 2022-04-06
US
Avanos Medical, Inc. recalls BALLARD* Multi-Access Port Catheter, Double Swivel Y Elbow, Adult, 14 French; Sterile; Single Use
Product was distributed with incorrect expiration date.
FDA · DeviceAvanos Medical, Inc.Worldwide distribution - US Nationwide distribution in the states of Oklahoma…Official notice →US
Microbiologics Inc recalls Pneumonia (33 Targets) Control Panel, unassayed quality control material, containing: Positive Control 1 - Semi-Quantitative Bacteria…
Shelf-life for the 8244 Pneumonia (33 Targets) Control Panel determined to be 6 months instead of 18 months.
FDA · DeviceMicrobiologics IncUS distribution to Alabama, Georgia, Illinois, Montana, Pennsylvania,…Official notice →US
Howmedica Osteonics Corp. recalls Biolox delta Ceramic V40 Femoral Head 32/ +4.0mm Catalog Number: 6570-0-232
Potential product mix where the size and/or offset of the Biolox delta Ceramic V40 Femoral Head inside the package does not match the package labeling
FDA · DeviceHowmedica Osteonics Corp.Worldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →US
Ansell Healthcare Products LLC recalls Gammex Non-Latex PI Green PI-KARE Skin-friendly PI Technology polyisoprene surgical gloves 7 1/2
Transport company failed to stop at the border for required FDA inspection.
FDA · DeviceAnsell Healthcare Products LLCU.S.: AZ, CA, NV, TX, UT and WA O.U.S.: NoneOfficial notice →US
ROi CPS LLC recalls Regard Custom Procedure Kits containing BD Luer Lok Access Devices
Recalled devices were packaged in convenience kits.
US
Meridian Bioscience Inc recalls Revogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories. Meridian branded Revogene: Catalog…
The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.
FDA · DeviceMeridian Bioscience IncWorldwide distribution - US Nationwide distribution in the states of AK, AL,…Official notice →