Drugs & medicalClass II

Abbott Laboratories, Inc recalls Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro…

Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.

Read the official notice →
Date
2022-03-23
Country
United States
Brand
Abbott Laboratories
Product
Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro…
Classification
Class II
Units
Worldwide distribution: US (nationwide) including states of: AL, CA, GA, IL,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Laboratories recalls

  1. 2022-08-03
  2. US
    Drugs & medicalClass II

    Abbott Laboratories recalls REF: 08P4377, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL

    There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.

    FDA · DeviceAbbott LaboratoriesUS: AK AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND…Official notice →
  3. 2022-06-29
  4. US
    Food & drinkClass II

    Abbott Laboratories recalls Pedialyte with Immune Support***+Zinc/Selenium & Magnesium***Peach Mango Natural & Artificial Flavor, 33.8 FL OZ (1.05 QT)…

    These products are being recalled due to the potential of leakage from the 1L container.

    FDA · FoodAbbott LaboratoriesAL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MI, MO, MS, NC, NH,…Official notice →
  5. 2022-06-22
  6. US
    Food & drinkClass III

    Abbott Laboratories recalls Pedialyte***Replaces Electrolytes***Great for Kids and Adults***Strawberry Lemonade Natural Flavor with Other Natural Flavor*** 6 0.6 OZ…

    The third party contract manufacturer used a dextrose ingredient derived from wheat in one lot of the Strawberry Lemonade powder, which is not declared on the label.

    FDA · FoodAbbott LaboratoriesAZ, CA, FL, GA, IL, IN, MN, NC, NY, OH, KS, and TN.Official notice →
  7. 2022-04-13
  8. US
    Drugs & medicalClass II

    Abbott Laboratories recalls ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT…

    This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).

    FDA · DeviceAbbott LaboratoriesWorldwide Distribution: US (nationwide) and OUS (foreign) including countries…Official notice →
  9. 2022-03-16
  10. US
    Food & drinkClass I

    Abbott Laboratories recalls Abbott Similac Advance Infant Formula with Iron 12.4oz. can, 23.2oz. plastic jar, 20.6oz. plastic jar. 6 units per case. Abbott Similac…

    Consumer complaints related to Cronobacter sakazakii or Salmonella Newport in infants and evidence of Cronobacter sakazakii in the manufacturing facility where the product had been produced

    FDA · FoodAbbott LaboratoriesUS Nationwide. Worldwide.Official notice →
  11. 2021-12-01
  12. US
    Drugs & medicalClass II

    Abbott Laboratories recalls ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89

    Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash…

    FDA · DeviceAbbott LaboratoriesU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →
All 21 Abbott Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2022-03-23
  2. US
    Drugs & medicalClass II

    Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls LINQ II Insertable Cardiac Monitor

    Devices may lose functionality due to susceptibility to moisture ingress.

    FDA · DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Worldwide distribution - US Nationwide distribution in the states of NJ, NY,…Official notice →
  3. US
    Drugs & medicalClass II

    Boston Scientific Corporation recalls HYDRATOME RX 44-20MM/450CM Material Number: M00583060

    Sterility of device is compromised due to a sterile barrier breach

    FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  4. US
    Drugs & medicalClass II

    Hologic, Inc recalls Panther Fusion Extraction Reagent-X, REF PRD-04477, For use with Panther Fusion System, IVD, Rx only, Contents: FCR-X Fusion Capture…

    Failed stability specifications for reagent kits that may result in false negative test results.

    FDA · DeviceHologic, IncWorldwide distribution - US Nationwide distribution in the states of AR, CA,…Official notice →
  5. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc. recalls Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system…

    The clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study…

    FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution: US (nationwide) and countries of: Algeria, Andorra,…Official notice →
  6. US
    Drugs & medicalClass I

    SD Biosensor, Inc. recalls STANDARD Q COVID-19 Ag Home Test STANDARD Q COVID-19 Ag Home Test is a rapid chromatographic immunoassay for the qualitative detection of…

    Illegal importation of COVID-19 AG Test kits into the U.S. The COVID-19 Ag Test are not approved, cleared or authorized for distribution within the U.S.

    FDA · DeviceSD Biosensor, Inc.U.S. Nationwide DistributionOfficial notice →
  7. US
    Drugs & medicalClass I

    Walmart Stores recalls Avaphinal Maximum Male Enhancement Sexual Pills, 2000 MG*, packaged in 10 Capsules per carton.

    Marketed Without An Approved NDA/ANDA: Product was found to contain undeclared sildenafil, an ingredient found in FDA approved products for the treatment of male sexual enhancement, making this an unapproved drug.

    FDA · DrugWalmart StoresNationwide in the USA via www.walmart.comOfficial notice →
See all drugs & medical recalls →