Abbott Laboratories recalls ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89
Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash…
Read the official notice →- Date
- 2021-12-01
- Country
- United States
- Brand
- Abbott Laboratories
- Product
- ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89
- Classification
- Class II
- Units
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Abbott Laboratories recalls
- 2022-08-03
US
Abbott Laboratories recalls REF: 08P4377, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
FDA · DeviceAbbott LaboratoriesUS: AK AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND…Official notice →- 2022-06-29
US
Abbott Laboratories recalls Pedialyte with Immune Support***+Zinc/Selenium & Magnesium***Peach Mango Natural & Artificial Flavor, 33.8 FL OZ (1.05 QT)…
These products are being recalled due to the potential of leakage from the 1L container.
FDA · FoodAbbott LaboratoriesAL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MI, MO, MS, NC, NH,…Official notice →- 2022-06-22
US
Abbott Laboratories recalls Pedialyte***Replaces Electrolytes***Great for Kids and Adults***Strawberry Lemonade Natural Flavor with Other Natural Flavor*** 6 0.6 OZ…
The third party contract manufacturer used a dextrose ingredient derived from wheat in one lot of the Strawberry Lemonade powder, which is not declared on the label.
- 2022-04-13
US
Abbott Laboratories recalls ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT…
This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).
FDA · DeviceAbbott LaboratoriesWorldwide Distribution: US (nationwide) and OUS (foreign) including countries…Official notice →- 2022-03-23
US
Abbott Laboratories, Inc recalls Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro…
Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.
FDA · DeviceAbbott Laboratories, IncWorldwide distribution: US (nationwide) including states of: AL, CA, GA, IL,…Official notice →- 2022-03-16
US
Abbott Laboratories recalls Abbott Similac Advance Infant Formula with Iron 12.4oz. can, 23.2oz. plastic jar, 20.6oz. plastic jar. 6 units per case. Abbott Similac…
Consumer complaints related to Cronobacter sakazakii or Salmonella Newport in infants and evidence of Cronobacter sakazakii in the manufacturing facility where the product had been produced
Other drugs & medical recalls in the United States
- 2021-12-01
US
Intuitive Surgical, Inc. recalls The da Vinci Xi and X surgical systems (IS4000 and IS4200) each consist of three main components: the Surgeon Console (SC), the Vision…
Incorrect expiration date on its package labeling (shipper box and inner carton label).
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, CA,…Official notice →US
The Binding Site Group, Ltd. recalls Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10)…
Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6…
FDA · DeviceThe Binding Site Group, Ltd.U.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: CanadaOfficial notice →US
Medtronic Perfusion Systems recalls Cardioblate CryoFlex Surgical Ablation Console
There is potential for intermittent electrical connectivity between the console and probe.
FDA · DeviceMedtronic Perfusion SystemsDistribution in US - IL, KS, and UT Distribution outside of US to Australia…Official notice →US
Western/Scott Fetzer Company recalls Western Oxytote DTE Digital VIPR System (with D-size cylinder) Model: MNDS-602
VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke
US
GE Healthcare, LLC recalls Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the…
US
Philips Healthcare recalls CombiDiagnost R90 is multi-functional general R/F systems.
The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and…