Drugs & medicalClass II

Medtronic Perfusion Systems recalls Cardioblate CryoFlex Surgical Ablation Console

There is potential for intermittent electrical connectivity between the console and probe.

Read the official notice →
Date
2021-12-01
Country
United States
Brand
Medtronic Perfusion Systems
Product
Cardioblate CryoFlex Surgical Ablation Console
Classification
Class II
Units
Distribution in US - IL, KS, and UT Distribution outside of US to Australia…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Medtronic Perfusion Systems recalls

  1. 2024-05-22
  2. US
  3. 2024-03-27
  4. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Medtronic DLP Vessel Cannula, Model Number REF 30000

    Medtronic is writing to inform you of incorrect labeling for three manufactured lots of the DLP Vessel Cannulae for the model and lot numbers listed above.

    FDA · DeviceMedtronic Perfusion SystemsUS and Brazil, Canada, Republic of Georgia, Germany, India, Ireland, Italy,…Official notice →
  5. 2023-02-08
  6. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Custom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF…

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or…

    FDA · DeviceMedtronic Perfusion SystemsWorldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia,…Official notice →
  7. 2022-06-01
  8. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the…

    Firm detected an increase in complaints related to fractured jaw tips of the Cardioblate Gemini-s.

    FDA · DeviceMedtronic Perfusion SystemsWorldwide distribution - US Nationwide distribution in the states of CA, ND,…Official notice →
  9. 2022-01-19
  10. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Medtronic Custom Perfusion kits, with the following product descriptions and corresponding model numbers: 1. CUSTOM PACK BB10Q85R…

    Product is labeled as non-pyrogenic but endotoxin testing was not performed.

    FDA · DeviceMedtronic Perfusion SystemsWorldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →
  11. 2021-10-13
  12. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Medtronic Bio-Medicus Insertion Kits, Model number 96553, Pediatric Insertion Kit, Sterile; and Model number 96551, Venous Insertion…

    There was a label mix-up between two model numbers resulting in the incorrect guidewire being contained in the kits for both model numbers.

    FDA · DeviceMedtronic Perfusion SystemsWorldwide distribution - US Nationwide distribution in the state of…Official notice →
All 21 Medtronic Perfusion Systems recalls →

Other drugs & medical recalls in the United States

  1. 2021-12-01
  2. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

    The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or…

    FDA · DeviceInstrumentation LaboratoryWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
  3. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls The da Vinci Xi and X surgical systems (IS4000 and IS4200) each consist of three main components: the Surgeon Console (SC), the Vision…

    Incorrect expiration date on its package labeling (shipper box and inner carton label).

    FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, CA,…Official notice →
  4. US
    Drugs & medicalClass II

    The Binding Site Group, Ltd. recalls Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10)…

    Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6…

    FDA · DeviceThe Binding Site Group, Ltd.U.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: CanadaOfficial notice →
  5. US
    Drugs & medicalClass II

    Abbott Laboratories recalls ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89

    Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash…

    FDA · DeviceAbbott LaboratoriesU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →
  6. US
    Drugs & medicalClass II

    Western/Scott Fetzer Company recalls Western Oxytote DTE Digital VIPR System (with D-size cylinder) Model: MNDS-602

    VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke

    FDA · DeviceWestern/Scott Fetzer CompanyCT, MD, MT, OHOfficial notice →
  7. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition

    The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the…

    FDA · DeviceGE Healthcare, LLCWorldwide and US Nationwide DistributionOfficial notice →
See all drugs & medical recalls →