Drugs & medicalClass II

NRT X-RAY A/S recalls Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system

X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75…

Read the official notice →
Date
2024-05-15
Country
United States
Brand
Nrt X-Ray A/S
Product
Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
Classification
Class II
Units
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Nrt X-Ray A/S recalls

  1. 2023-06-07
  2. US
    Drugs & medicalClass II

    NRT X-RAY A/S recalls Intelli-C, Left side suspended Tabletop, REF: 03400010

    There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

    FDA · DeviceNRT X-RAY A/SUS Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA,…Official notice →
  3. 2022-02-02
  4. US
    Drugs & medicalClass II

    NRT X-RAY A/S recalls 1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C…

    Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C,…

    FDA · DeviceNRT X-RAY A/SUSA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic,…Official notice →
All 3 Nrt X-Ray A/S recalls →

Other drugs & medical recalls in the United States

  1. 2024-05-22
  2. US
    Drugs & medicalClass I

    Stop Clopez Corp recalls Schwinnng Herbal Dietary Supplement Capsules, packaged in 10-count boxes, distributed by VSD Productions Inc., Las Vegas, NV.

    Marketed Without An Approved NDA/ANDA

    FDA · DrugStop Clopez CorpDistribution worldwide via amazon.comOfficial notice →
  3. 2024-05-15
  4. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of…

    Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Stryker Leibinger GmbH & Co. KG recalls Stryker 1,5/1,7MM DYNAMIC MESH- STD-MEDIUM-Intended for the reconstruction, stabilization and rigid fixation of the bony areas…

    Mislabeled: Product labeled 56-90614 (90x90 Gold Mesh), was actually 56-90314 (90x90 Blue Mesh), may cause a delay in procedure

    FDA · DeviceStryker Leibinger GmbH & Co. KGNationwideOfficial notice →
  6. US
    Drugs & medicalClass II

    GE Medical Systems, SCS recalls GE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systems

    GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could…

    FDA · DeviceGE Medical Systems, SCSWorldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA,…Official notice →
  7. US
    Drugs & medicalClass I

    Medtronic Navigation, Inc. recalls Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating…

    Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.

    FDA · DeviceMedtronic Navigation, Inc.Worldwide distribution. US nationwide including Puerto Rico, Argentina,…Official notice →
  8. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Baxter Spectrum IQ Infusion Pump, Product Code 3570009

    One device was improperly performed testing prior to release from a Service Center.

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the state of Massachusetts.Official notice →
See all drugs & medical recalls →