Medtronic Navigation, Inc. recalls Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating…
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
Read the official notice →- Date
- 2024-05-15
- Country
- United States
- Brand
- Medtronic Navigation
- Product
- Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating…
- Classification
- Class I
- Units
- Worldwide distribution. US nationwide including Puerto Rico, Argentina,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Medtronic Navigation recalls
- 2025-02-05
US
Medtronic Navigation, Inc. recalls Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The Sterile…
Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.
FDA · DeviceMedtronic Navigation, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →- 2024-10-16
US
Medtronic Navigation, Inc. recalls StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is…
Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing…
FDA · DeviceMedtronic Navigation, Inc.Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ,…Official notice →- 2024-01-17
US
Medtronic Navigation, Inc. recalls Medtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUS
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
FDA · DeviceMedtronic Navigation, Inc.Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA and the…Official notice →- 2023-11-15
US
Medtronic Navigation, Inc. recalls StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation Cranial software, is…
Due to a software anomaly which potentially could result in the surgical planning data shifting to an unintended location.
FDA · DeviceMedtronic Navigation, Inc.Worldwide Distribution: U.S. (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE,…Official notice →- 2023-06-14
US
Medtronic Navigation, Inc. recalls StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems
In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain…
FDA · DeviceMedtronic Navigation, Inc.US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA,…Official notice →US
Medtronic Navigation, Inc. recalls StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems
During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information…
FDA · DeviceMedtronic Navigation, Inc.Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR,…Official notice →
Other drugs & medical recalls in the United States
- 2024-05-15
US
Stryker Leibinger GmbH & Co. KG recalls Stryker 1,5/1,7MM DYNAMIC MESH- STD-MEDIUM-Intended for the reconstruction, stabilization and rigid fixation of the bony areas…
Mislabeled: Product labeled 56-90614 (90x90 Gold Mesh), was actually 56-90314 (90x90 Blue Mesh), may cause a delay in procedure
US
NRT X-RAY A/S recalls Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75…
FDA · DeviceNRT X-RAY A/SUS Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV,…Official notice →US
GE Medical Systems, SCS recalls GE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systems
GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could…
FDA · DeviceGE Medical Systems, SCSWorldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA,…Official notice →US
Baxter Healthcare Corporation recalls Baxter Spectrum IQ Infusion Pump, Product Code 3570009
One device was improperly performed testing prior to release from a Service Center.
FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the state of Massachusetts.Official notice →US
US
Angiodynamics, Inc. recalls DURAFLOW 2, 32CM STRAIGHT BASIC KIT PG-Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product…
Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.