  1. 2025-02-05
  2. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The Sterile…

    Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.

    FDA · DeviceMedtronic Navigation, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  3. 2024-10-16
  4. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is…

    Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing…

    FDA · DeviceMedtronic Navigation, Inc.Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ,…Official notice →
  5. 2024-05-15
  6. US
    Drugs & medicalClass I

    Medtronic Navigation, Inc. recalls Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating…

    Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.

    FDA · DeviceMedtronic Navigation, Inc.Worldwide distribution. US nationwide including Puerto Rico, Argentina,…Official notice →
  7. 2024-01-17
  8. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls Medtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUS

    Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.

    FDA · DeviceMedtronic Navigation, Inc.Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA and the…Official notice →
  9. 2023-11-15
  10. US
    Drugs & medicalClass I

    Medtronic Navigation, Inc. recalls StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation Cranial software, is…

    Due to a software anomaly which potentially could result in the surgical planning data shifting to an unintended location.

    FDA · DeviceMedtronic Navigation, Inc.Worldwide Distribution: U.S. (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE,…Official notice →
  11. 2023-06-14
  12. US
    Drugs & medicalClass I

    Medtronic Navigation, Inc. recalls StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems

    In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain…

    FDA · DeviceMedtronic Navigation, Inc.US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA,…Official notice →
  13. US
    Drugs & medicalClass I

    Medtronic Navigation, Inc. recalls StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems

    During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information…

    FDA · DeviceMedtronic Navigation, Inc.Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR,…Official notice →
  14. 2023-02-08
  15. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls AxiEM" Non-Invasive Patient Tracker

    Due to increasing complaint trend for the Non-Invasive Patient Tracker system where complaints analysis indicates that users are unable to successfully verify their navigation instruments during image-guided surgery.

    FDA · DeviceMedtronic Navigation, Inc.Worldwide - U.S. distribution including in the states of AL, AR, AZ, CA, CO,…Official notice →
  16. 2023-01-18
  17. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls RadiaLux Lighted Retractor (Pink), REF: 50-101-1

    Defect in sterile pouch seal, incomplete seal, of lighted Retractors, increases the risk of contamination, which can lead to post-operative infection.

    FDA · DeviceMedtronic Navigation, Inc.US Nationwide distribution including in the states of OH, CA, FL, NJ, NY, MN,…Official notice →
  18. 2022-11-23
  19. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necrotize or coagulate soft tissue…

    Due to a defect in the outer pouch sterile seal

    FDA · DeviceMedtronic Navigation, Inc.U.S. Nationwide distribution including in the states AZ, CA, CO, DC, FL, IL,…Official notice →
  20. 2021-12-29
  21. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the…

    A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy…

    FDA · DeviceMedtronic Navigation, Inc.US: FL, NE, CA, MN, TX, OH, IL, AZ, KS, KY, OR, OK, WA, NV, MO, IN, MI, UT,…Official notice →
  22. US
    Drugs & medicalClass I

    Medtronic Navigation, Inc. recalls Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model:…

    Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy…

    FDA · DeviceMedtronic Navigation, Inc.US: MN, FL, MD, IL, AK, NE, NH, PA, ND, AR, CA, WI, NJ, GA, SD, AZ, KY, AL,…Official notice →