NRT X-RAY A/S recalls Intelli-C, Left side suspended Tabletop, REF: 03400010
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Read the official notice →- Date
- 2023-06-07
- Country
- United States
- Brand
- Nrt X-Ray A/S
- Product
- Intelli-C, Left side suspended Tabletop, REF: 03400010
- Classification
- Class II
- Units
- US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Nrt X-Ray A/S recalls
- 2024-05-15
US
NRT X-RAY A/S recalls Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75…
FDA · DeviceNRT X-RAY A/SUS Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV,…Official notice →- 2022-02-02
US
NRT X-RAY A/S recalls 1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C…
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C,…
FDA · DeviceNRT X-RAY A/SUSA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic,…Official notice →
Other drugs & medical recalls in the United States
- 2023-06-14
US
Empower Clinic Services LLC dba Empower Pharmacy recalls ASCORBIC ACID PRESERVED INJECTION SOLUTION, 500 mg/mL, 30 mL Sterile Multiple-Dose Vial, RX ONLY, Compounded…
Mislabeling: preservative free product labeled as preserved.
FDA · DrugEmpower Clinic Services LLC dba Empower PharmacyDistributed Nationwide in the USAOfficial notice →- 2023-06-07
US
Linkbio Corp. recalls LINK SymphoKnee System L-Shaped Femoral Augment, Medial-Right/Lateral-Left, Sz 5-6 Tilastan, cemented. Intended for primary and revision…
Contains the incorrect implant in the packaging. The packaging incorrectly contains the opposite side of the implant (Lateral-Right/Medial-Left instead of Medial-Right/Lateral-Left).
US
Smith & Nephew, Inc. recalls Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200…
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an…
US
Elekta, Inc. recalls Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy
A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601
FDA · DeviceElekta, Inc.Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA.…Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 11 VANC Reagent Generation (GEN) 50 with affected Assay Data Disk (ADD) for Data…
Unavailable to calibrate VITROS Chemistry Products VANC Reagent, Generation (GEN) 50, results in delay of testing
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide Foreign: Brazil, Chile, ColombiaOfficial notice →US
TELEFLEX LLC recalls Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or…
Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step…