Drugs & medicalClass II

MICROVENTION INC. recalls LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm,…

Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress,…

Read the official notice →
Date
2026-05-20
Country
United States
Brand
Microvention
Product
LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm,…
Classification
Class II
Units
International distribution to the country of China.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Microvention recalls

  1. 2026-08-19
  2. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls BOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in facilitating the insertion and…

    Due to an increase rate of balloon pinhole defects.

    FDA · DeviceMICROVENTION INC.Worldwide - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →
  3. 2025-10-22
  4. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536,…

    Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may…

    FDA · DeviceMICROVENTION INC.Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL,…Official notice →
  5. 2025-08-06
  6. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400

    Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional…

    FDA · DeviceMICROVENTION INC.OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France,…Official notice →
  7. 2025-02-19
  8. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended…

    Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.

    FDA · DeviceMICROVENTION INC.US: None OUS: ChinaOfficial notice →
  9. 2025-01-29
  10. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Sterile R, Rx Only, Lot# 0000629846

    Sterility assurance; microcatheter inner packaging may contain incomplete seal.

    FDA · DeviceMICROVENTION INC.US: None OUS: ChinaOfficial notice →
  11. 2024-09-25
  12. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls ERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomy device designed to restore…

    Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.

    FDA · DeviceMICROVENTION INC.U.S. Nationwide distribution in the states of AR, CA, FL, IN, LA, MA, MN, MO,…Official notice →
All 12 Microvention recalls →

Other drugs & medical recalls in the United States

  1. 2026-05-27
  2. US
    Drugs & medicalClass III

    Golden State Medical Supply Inc. recalls Primidone Tablets, USP, 50 mg, 50-count bottle, RX only, Manufactured by: Lannett Company, Inc., Marketed by: GSMS,…

    Cross contamination with other products: API contaminated with trace amounts of Acemetacin API.

    FDA · DrugGolden State Medical Supply Inc.USA NationwideOfficial notice →
  3. US
    Drugs & medicalClass III

    Lannett Company Inc. recalls Primidone Tablets USP, 250mg, 100-count bottle, Rx only, Distributed by: Lannett Company, Inc., Philadelphia, PA 19136, NDC 0527-1231-01.

    Cross contamination with other products: API contaminated with trace amounts of Acemetacin API.

    FDA · DrugLannett Company Inc.USA NationwideOfficial notice →
  4. 2026-05-20
  5. US
    Drugs & medicalClass II

    Medtronic Neuromodulation recalls Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG…

    Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare…

    FDA · DeviceMedtronic NeuromodulationWorldwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Stryker Sustainability Solutions recalls BARD Dynamic Tip Steerable, Product Number 6DYNTP001;

    Incomplete seals on sterile product

    FDA · DeviceStryker Sustainability SolutionsUS Nationwide and the countries of Israel and Canada.Official notice →
  7. US
    Drugs & medicalClass II

    Integra LifeSciences Corp. (NeuroSciences) recalls Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD…

    Out-of-specification endotoxin result that did not meet the acceptance criteria.

    FDA · DeviceIntegra LifeSciences Corp. (NeuroSciences)US Nationwide distribution in the states of Arizona (AZ), Arkansas (AR),…Official notice →
  8. US
    Drugs & medicalClass II

    Edwards Lifesciences, LLC recalls Edwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EO

    Labeling update to provide warning if functionality of the valve replacement delivery system is compromised.

    FDA · DeviceEdwards Lifesciences, LLCWorldwide - US Nationwide distribution including in the states of AR, AZ, CA,…Official notice →
See all drugs & medical recalls →