Drugs & medicalClass II

MICROVENTION INC. recalls LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400

Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional…

Read the official notice →
Date
2025-08-06
Country
United States
Brand
Microvention
Product
LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400
Classification
Class II
Units
OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Microvention recalls

  1. 2026-08-19
  2. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls BOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in facilitating the insertion and…

    Due to an increase rate of balloon pinhole defects.

    FDA · DeviceMICROVENTION INC.Worldwide - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →
  3. 2026-05-20
  4. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm,…

    Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress,…

    FDA · DeviceMICROVENTION INC.International distribution to the country of China.Official notice →
  5. 2025-10-22
  6. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536,…

    Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may…

    FDA · DeviceMICROVENTION INC.Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL,…Official notice →
  7. 2025-02-19
  8. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended…

    Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.

    FDA · DeviceMICROVENTION INC.US: None OUS: ChinaOfficial notice →
  9. 2025-01-29
  10. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Sterile R, Rx Only, Lot# 0000629846

    Sterility assurance; microcatheter inner packaging may contain incomplete seal.

    FDA · DeviceMICROVENTION INC.US: None OUS: ChinaOfficial notice →
  11. 2024-09-25
  12. US
    Drugs & medicalClass II

    MICROVENTION INC. recalls ERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomy device designed to restore…

    Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.

    FDA · DeviceMICROVENTION INC.U.S. Nationwide distribution in the states of AR, CA, FL, IN, LA, MA, MN, MO,…Official notice →
All 12 Microvention recalls →

Other drugs & medical recalls in the United States

  1. 2025-08-06
  2. US
    Drugs & medicalClass II

    Phasor Health, LLC recalls Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35

    Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.

    FDA · DevicePhasor Health, LLCUS: PA, IN, AZ, MO, LA, FL, TX, MA, OH, TN, MI, WV, VA, CA, DE, NH, GA, SC,…Official notice →
  3. US
    Drugs & medicalClass II

    Boston Foundation For Sight recalls BostonSight SCLERAL Lens

    Select orders of lenses were manufactured with errors impacting sagittal height, haptics, or both.

    FDA · DeviceBoston Foundation For SightWorldwide distribution - US Nationwide and the countries of Canada.Official notice →
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura R8.2.x Systems. Software Verion: R8.2.x. 1. Allura Xper FD10, Model Number: 722026. 2. Allura Xper…

    Potential for temporary loss of imaging (X-ray) functionality due to software issue.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide and the countries of Algeria, Argentina, Australia,…Official notice →
  5. US
    Drugs & medicalClass II

    Steris Corporation recalls Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

    The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring…

    FDA · DeviceSteris CorporationWorldwide - US Nationwide and Puerto Rico and the country of Canada.Official notice →
  6. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an…

    Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead…

    FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and…Official notice →
  7. US
    Drugs & medicalClass II

    HeartSine Technologies Ltd recalls HeartSine SAM 350P, HeartSine SAM 360P, and HeartSine SAM 450P The HeartSine samaritan¿ PAD (Public Access Defibrillator)

    Due to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks)

    FDA · DeviceHeartSine Technologies LtdWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
See all drugs & medical recalls →