MICROVENTION INC. recalls Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536,…
Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may…
Read the official notice →- Date
- 2025-10-22
- Country
- United States
- Brand
- Microvention
- Product
- Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536,…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Microvention recalls
- 2026-08-19
US
MICROVENTION INC. recalls BOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in facilitating the insertion and…
Due to an increase rate of balloon pinhole defects.
FDA · DeviceMICROVENTION INC.Worldwide - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →- 2026-05-20
US
MICROVENTION INC. recalls LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm,…
Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress,…
- 2025-08-06
US
MICROVENTION INC. recalls LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400
Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional…
FDA · DeviceMICROVENTION INC.OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France,…Official notice →- 2025-02-19
US
MICROVENTION INC. recalls MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended…
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
- 2025-01-29
US
MICROVENTION INC. recalls Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Sterile R, Rx Only, Lot# 0000629846
Sterility assurance; microcatheter inner packaging may contain incomplete seal.
- 2024-09-25
US
MICROVENTION INC. recalls ERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomy device designed to restore…
Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.
FDA · DeviceMICROVENTION INC.U.S. Nationwide distribution in the states of AR, CA, FL, IN, LA, MA, MN, MO,…Official notice →
Other drugs & medical recalls in the United States
- 2025-10-22
US
Insulet Corporation recalls The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field…
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a…
US
Contego Medical LLC recalls Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF:…
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The…
FDA · DeviceContego Medical LLCUS Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA,…Official notice →US
Abbott recalls TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheter
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
US
Boston Scientific Corporation recalls Model Number L211 PROPONENT DR SL MRI Pacemaker
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes…
US
Olympus Corporation of the Americas recalls EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE XT160
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
US
Cook Medical Incorporated recalls Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W…
Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization.
FDA · DeviceCook Medical IncorporatedDomestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC…Official notice →