Microgenics Corporation recalls QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL…
Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
Read the official notice →- Date
- 2022-07-06
- Country
- United States
- Brand
- Microgenics
- Product
- QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL…
- Classification
- Class II
- Units
- U.S.: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Microgenics recalls
- 2025-11-05
US
Microgenics Corporation recalls Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303,…
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.
FDA · DeviceMicrogenics CorporationWorldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI,…Official notice →
Other drugs & medical recalls in the United States
- 2022-07-06
US
Baxter Healthcare Corporation recalls #1 MAC F/O LARYNGOSCOPE; F/O Laryngoscope Set-MAC; F/O Laryngoscope Set-MAC W/LED
Packaging error.
FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the countries of Japan, Canada,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline Pill Splitters; 1 each/box; 144 boxes/carton
Product lid (which houses a blade) is coming unhinged from the base, thus exposing the blade.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Singapore,…Official notice →US
GE Healthcare, LLC recalls Back up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019
Backup batteries can fail earlier than their estimated life.
FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →US
Akorn, Inc. recalls Fluticasone Propionate Nasal Spray, USP, 50 mcg, 16 g net fill weight per amber glass bottle, Rx only, Manufactured by: HI-TECH PHARMACAL…
Defective container: defect prevents product from dispensing as intended.
US
US
Akorn, Inc. recalls Calcipotriene Topical Solution, 0.005% (Scalp Solution), Rx only, 60 mL (2 fl. oz.) bottle, Manufactured by: HI-TECH PHARMACAL CO., INC.,…
Defective Delivery System: Potential defect that could prevent the product from dispensing as intended.