GE Healthcare, LLC recalls Back up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019
Backup batteries can fail earlier than their estimated life.
Read the official notice →- Date
- 2022-07-06
- Country
- United States
- Brand
- GE Healthcare
- Product
- Back up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Algeria,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More GE Healthcare recalls
- 2022-07-27
US
GE Healthcare, LLC recalls GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000.
The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.
FDA · DeviceGE Healthcare, LLCDistribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD,…Official notice →US
GE Healthcare, LLC recalls GE Healthcare Aisys CS2 Anesthesia System, Part number 1011-9050-000; and Avance CS2 and Avance CS2 Pro Anesthesia System, Part number…
There is a potential reversal of the O2 and air cylinder pressure transducer connections.
FDA · DeviceGE Healthcare, LLCWorldwide Distribution. US nationwide, Australia, Austria, Belgium, Bolivia,…Official notice →- 2022-07-13
US
GE Healthcare, LLC recalls GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data within a computer…
Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)
- 2022-06-22
US
GE Healthcare, LLC recalls CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part…
The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.
FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide. There was also military/government…Official notice →- 2022-06-15
US
GE Healthcare, LLC recalls Centricity Radiology RA600
Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values…
FDA · DeviceGE Healthcare, LLCWorldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC,…Official notice →US
GE Healthcare, LLC recalls Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstr¿om Carestation and Engstr¿om PRO…
Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.
FDA · DeviceGE Healthcare, LLCWorldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI,…Official notice →
Other drugs & medical recalls in the United States
- 2022-07-06
US
Canon Medical System, USA, INC. recalls System: Alphenix 4D CT in combination with CAS-930A (C arm Support) Model Number: INFX-8000C/BF, INFX-8000C/BH, INFX-8000C/SW
CT operation may be restricted by an interlock which is a result of an system error and the CT system may stop operating properly after an attempt to cancel the error message has been initiated.
FDA · DeviceCanon Medical System, USA, INC.U.S.: CA, CO, FL, NY, and WY O.U.S.: NoneOfficial notice →US
Baxter Healthcare Corporation recalls #1 MAC F/O LARYNGOSCOPE; F/O Laryngoscope Set-MAC; F/O Laryngoscope Set-MAC W/LED
Packaging error.
FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the countries of Japan, Canada,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline Pill Splitters; 1 each/box; 144 boxes/carton
Product lid (which houses a blade) is coming unhinged from the base, thus exposing the blade.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Singapore,…Official notice →US
Microgenics Corporation recalls QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL…
Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
FDA · DeviceMicrogenics CorporationU.S.: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC,…Official notice →US
Akorn, Inc. recalls Fluticasone Propionate Nasal Spray, USP, 50 mcg, 16 g net fill weight per amber glass bottle, Rx only, Manufactured by: HI-TECH PHARMACAL…
Defective container: defect prevents product from dispensing as intended.
US